Purpose: To develop a free-breathing myocardial mapping technique using inversion-recovery (IR) radial fast low-angle shot (FLASH) and calibrationless motion-resolved model-based reconstruction.

Methods: Free-running (free-breathing, retrospective cardiac gating) IR radial FLASH is used for data acquisition at 3T. First, to reduce the waiting time between inversions, an analytical formula is derived that takes the incomplete recovery into account for an accurate calculation. Second, the respiratory motion signal is estimated from the k-space center of the contrast varying acquisition using an adapted singular spectrum analysis (SSA-FARY) technique. Third, a motion-resolved model-based reconstruction is used to estimate both parameter and coil sensitivity maps directly from the sorted k-space data. Thus, spatiotemporal total variation, in addition to the spatial sparsity constraints, can be directly applied to the parameter maps. Validations are performed on an experimental phantom, 11 human subjects, and a young landrace pig with myocardial infarction.

Results: In comparison to an IR spin-echo reference, phantom results confirm good accuracy, when reducing the waiting time from 5 s to 1 s using the new correction. The motion-resolved model-based reconstruction further improves precision compared to the spatial regularization-only reconstruction. Aside from showing that a reliable respiratory motion signal can be estimated using modified SSA-FARY, in vivo studies demonstrate that dynamic myocardial maps can be obtained within 2 min with good precision and repeatability.

Conclusion: Motion-resolved myocardial mapping during free-breathing with good accuracy, precision and repeatability can be achieved by combining inversion-recovery radial FLASH, self-gating and a calibrationless motion-resolved model-based reconstruction.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9892313PMC
http://dx.doi.org/10.1002/mrm.29521DOI Listing

Publication Analysis

Top Keywords

motion-resolved model-based
20
model-based reconstruction
16
myocardial mapping
12
inversion-recovery radial
12
radial flash
12
free-breathing myocardial
8
calibrationless motion-resolved
8
waiting time
8
respiratory motion
8
motion signal
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!