Statement Of Problem: Screw loosening is the most common mechanical complication with implant prostheses. How the alteration of implant-to-abutment connection surfaces that occurs during laboratory procedures affects screw loosening is unclear.
Purpose: The purpose of this in vitro study was to compare the reverse tightening value (RTV) differences between custom castable abutments before casting, after casting in a conventional manner, and after casting with custom protector caps and pegs.
Material And Methods: Thirty implants with a standard-diameter conical connection (NobelReplace Conical Connection 4.3×13 mm; Nobel Biocare AG) and 30 premachined 4.3-mm GoldAdapt abutments (GoldAdapt; Nobel Biocare AG) were selected for this study. Specimens were divided into 3 groups (n=10): the uncast custom castable abutment group (UCCA) in which abutments were new and not cast; the unprotected custom castable abutment group (UPCCA) in which abutments were cast and devested with airborne-particle abrasion; and the protected custom castable abutment group (PCCA) in which abutments were cast by using protector caps and pegs made by milling zirconia and then devested with airborne-particle abrasion. All abutments in each group were tightened to 35 Ncm with a calibrated digital tightening device. After 10 minutes, all screws were retightened to 35 Ncm. At 3 hours, each screw was loosened, and the value at which the initial loosening occurred was documented as the RTV. The results were statistically analyzed with 1-way ANOVA to explore differences, and post hoc tests with Tukey adjustment were used for multiple comparisons.
Results: Among the tested groups, the mean RTV ranged from 19.89 Ncm to 27.19 Ncm: UCCA 27.19 Ncm, UPCCA 19.89 Ncm, and PCCA 24.24 Ncm. A significant difference was found among the tested groups (P<.05).
Conclusions: Casting procedures, especially devestment with airborne-particle abrasion, affected implant-abutment connections and the seat site of the screw. Protecting the implant connection site and the seat site of the abutment screw with protector caps and pegs prevented a significant loss of the RTV.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.prosdent.2022.09.006 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!