Objective: Currently, minimally invasive internal fixation is recommended for the surgical treatment of unstable pelvic fractures. The premise and difficulty of minimally invasive internal fixation are minimally invasive reduction of fractures. This review aimed to investigate the indications, surgical strategy and techniques, safety, and efficacy of intelligent robot-assisted fracture reduction (RAFR) system of pelvic ring injuries.
Methods: This retrospective study reviewed a case series from March 2021 to November 2021. A total of 22 patients with unstable pelvic fracture injuries underwent minimally invasive internal fixations. All pelvic ring fractures were reduced with our intelligent RAFR system. The robot system intelligently designs the optimal position and reduction path based on the patient's preoperative 3D CT. During the operation, the three-dimensional visualization of the fracture is realized through image registration, and the Robot completes the automatic reduction of the fracture. The global 3D point cloud error between the preoperative planning results and the actual postoperative reduction results was calculated. The postoperative reduction results of residual displacement were graded by the Matta Criteria.
Results: Minimally invasive closed reduction procedures were completed in all 22 cases with our RAFR system. The average global 3D point cloud reduction error between the preoperative planning results and the actual postoperative reduction results was 3.41mm±1.83mm. The mean residual displacement was 4.61mm±3.29mm. Given the Matta criteria, 16 cases were excellent, five were good, and one was fair, with an excellent and good rate of 95.5%.
Conclusion: Our new pelvic fracture reduction robot system can complete intelligent and minimally invasive fracture reduction for most patients with unstable pelvic fractures. The system has intelligent reduction position and path planning and realizes stable pelvis control through a unique holding arm and a robotic arm. The operation process will not cause additional damage to the patient, which fully meets the clinical requirements. Our study demonstrated the safety and effectiveness of our robotic reduction system and its applicability and usability in clinical practice, thus paving the way towards Robot minimally invasive pelvic fracture surgeries.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.injury.2022.11.001 | DOI Listing |
J Cosmet Dermatol
January 2025
Department of Dermatology, The Second Affiliated Hospital of Jiaxing University, Jiaxing, Zhejiang, People's Republic of China.
Objective: A simple and minimally invasive combined procedure, including transconjunctival orbital fat removal and transcutaneous resected orbital fat injection, was performed based on the anatomical characteristics of the lower eyelids in our young Chinese patients. Our study aimed to investigate the efficacy and safety of this procedure in our study population.
Methods: In our retrospective study, a total of 183 consecutive patients underwent a combination of traditional transconjunctival blepharoplasty and nanofat grafting between February 2020 and June 2024.
Rev Med Suisse
January 2025
Service d'oto-rhino-laryngologie et de chirurgie cervico-faciale, Centre hospitalier universitaire vaudois, 1011 Lausanne.
This article highlights recent advancements in otorhinolaryngology. It reviews innovations in managing sudden sensory-neural hearing loss through low-dose glucocorticoid treatments and explores pediatric audiology breakthroughs, particularly gene therapy for DFNB9 deafness. In speech pathology and oncology, a new minimally invasive robotic surgical approach enhances organ preservation for head and neck cancers.
View Article and Find Full Text PDFHSS J
February 2025
Department of Spine Surgery, Hospital for Special Surgery, New York, NY, USA.
The scope of existing annular closure device (ACD) studies examining long-term follow-up data is limited. There is a paucity of studies that report and analyze recent outcomes data following ACD use. We sought to summarize the available long-term follow-up data on postoperative outcomes of the Barricaid (Intrinsic Therapeutics) ACD.
View Article and Find Full Text PDFBackground And Aims: Pectus carinatum (PC) is the second most common deformity of the anterior chest wall, resulting in detrimental effects on body image and quality of life. This study evaluated the safety, effectiveness, and factors associated with the treatment of PC using a sandwiched bar and screw fixation system, first performed in Vietnam at the University Medical Center Ho Chi Minh City in 2016.
Methods: This retrospective cohort study was conducted from March 2016 to February 2023 in patients with PC and PC-mixed pectus excavatum (PE) deformities.
Trauma Surg Acute Care Open
January 2025
Department of Surgery, University of Colorado School of Medicine, Aurora, Colorado, USA.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!