Reproducibility of T and T quantification in a multi-vendor multi-site study.

Osteoarthritis Cartilage

Program of Advanced Musculoskeletal Imaging (PAMI), Cleveland Clinic, OH, USA; Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, OH, USA; Department of Diagnostic Radiology, Imaging Institute, Cleveland Clinic, OH, USA. Electronic address:

Published: February 2023

AI Article Synopsis

Article Abstract

Objective: To evaluate the multi-vendor multi-site reproducibility of two-dimensional (2D) multi-echo spin-echo (MESE) T mapping (product sequences); and to evaluate the longitudinal reproducibility of three-dimensional (3D) magnetization-prepared angle-modulated partitioned k-space spoiled gradient echo snapshots (MAPSS) T and T mapping (research sequences), and 2D MESE T mapping, separated by 6 months, in a multi-vendor multi-site setting.

Methods: Phantoms and volunteers (n = 5 from each site, n = 20 in total) were scanned on four 3 T magnetic resonance (MR) systems from four sites and three vendors (Siemens, General Electric, and Phillips). Two traveling volunteers (3 knees) scanned at all 4 sites at baseline and 6-month follow-up. Data was transferred to one site for centralized processing. Coefficients of variation (CVs) were calculated to evaluate reproducibility.

Results: For baseline 2D MESE T measures, average CV were 0.37-2.45% (intra-site) and 5.96% (inter-site) for phantoms, and 3.15-8.49% (intra-site) and 14.16% (inter-site) for volunteers. For longitudinal phantom data, intra-site CVs were 1.42-3.48% for 3D MAPSS T, 1.77-3.56% for 3D MAPSS T, and 1.02-2.54% for 2D MESE T. For the longitudinal volunteer data, the intra-site CVs were 2.60-4.86% for 3D MAPSS T, 3.33-7.25% for 3D MAPSS T, and 3.11-8.77% for 2D MESE T.

Conclusion: This study demonstrated excellent intra-site reproducibility of 2D MESE T imaging, while its inter-site variation was slightly higher than 3D MAPSS T imaging (10.06% as previously reported). This study also showed excellent reproducibility of longitudinal T and T cartilage quantification, in a multi-vendor multi-site setting for both product 2D MESE T and 3D MAPSS T/T research sequences.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10016129PMC
http://dx.doi.org/10.1016/j.joca.2022.10.017DOI Listing

Publication Analysis

Top Keywords

multi-vendor multi-site
16
quantification multi-vendor
8
mese mapping
8
data intra-site
8
intra-site cvs
8
mese
7
mapss
7
reproducibility
5
intra-site
5
reproducibility quantification
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!