To evaluate the stability of a dental implant and the effectiveness of a newly designed damping capacity assessment device by improving the number of blows and strength evaluated by a prospective clinical study. The stability of dental implants was measured in 50 implants in a total of 38 patients. Measurements were performed using Anycheck and Periotest M devices, twice in total, divided into buccal and lingual directions. In addition, measurements were performed on the day of surgery, two weeks, one month, two months, and three months after surgery for a total of five times. After the standardization of the measured values, the differences and changes over time for each device were observed. No difference in standardized values between the two devices was observed at any time point. In both devices, stability decreased at two weeks postoperatively but gradually increased thereafter. No differences were observed in the values according to the measurement direction. The damping capacity of Anycheck was similar to that of Periotest M. After a slight decrease in stability two weeks after implant placement, implant stability increased over time.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9695454PMC
http://dx.doi.org/10.3390/medicina58111570DOI Listing

Publication Analysis

Top Keywords

damping capacity
12
dental implant
8
implant stability
8
newly designed
8
designed damping
8
capacity assessment
8
assessment device
8
stability dental
8
measurements performed
8
anycheck periotest
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!