The instability of human serum albumin (HSA) in urine samples makes fresh urine a requirement for microalbumin analyses using immunoturbidimetry. Here, we determined the ability of an aptasensor-based fluorescent platform to detect microalbumin in old, boric acid-preserved urine samples. Our results show that the cleavage site of protease enzymes on urine albumin protein differed from the binding position of the aptamer on HSA protein, suggesting the aptasensor may be effective for albumin detection in non-fresh urine. Furthermore, the addition of boric acid in urine samples over a short term (at ambient temperature (T) and 4 °C), long term (-20 and -80 °C), and following freeze-thawing (1-3 cycles) did not significantly affect albumin stability, as analyzed using the aptasensor. Therefore, boric acid stabilized has in urine stored over a short- and long-term. Thus, the aptasensor developed by us is applicable for HSA detection in boric acid-preserved urine that has been stored for 7-d at T and 4 °C, and in the long-term at -80 °C.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9687920PMC
http://dx.doi.org/10.3390/bios12110915DOI Listing

Publication Analysis

Top Keywords

boric acid
12
urine samples
12
aptasensor-based fluorescent
8
urine
8
boric acid-preserved
8
acid-preserved urine
8
-80 °c
8
urine stored
8
effects boric
4
acid storage
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!