Purpose: Describe an early-adopting institution's experience with online adaptive radiation for stereotactic partial breast irradiation.

Methods And Materials: Retrospective review of 22 women treated between May 2021 and March 2022 with adaptive stereotactic partial breast irradiation. A total of 106 of 110 fractions were evaluated for dosimetric changes in target coverage and organ-at-risk (OAR) dose. Patient set up with stereotactic wooden frame and adapted per fraction. Treatment and planning times were collected prospectively by radiation therapists.

Results: Scheduled PTV was <95% in 72.1% and <90% in 38.5% of fractions, and both PTV and CTV coverage were improved significantly after adaption, and 83.7% of fractions were delivered as adapted per physician choice. There was no difference in OAR coverage. Average adaptive treatment planning took 15 min and average time-on-couch was 34.4 min.

Conclusions: Adaptive stereotactic breast irradiation resulted in improved target coverage with equivalent dosing to OARs in an efficient and tolerated treatment time. Improved target coverage allowed for decreased PTV margins compared to prior trial protocols that may improve acute and late toxicities.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9924103PMC
http://dx.doi.org/10.1002/acm2.13813DOI Listing

Publication Analysis

Top Keywords

stereotactic partial
12
partial breast
12
target coverage
8
online adaptive
8
adaptive stereotactic
8
breast irradiation
8
feasibility improved
4
improved target
4
coverage compromising
4
compromising organs
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!