Background: Our study aims to validate a commercially available inertial measurement unit (IMU) system against a standard laboratory-based optical motion capture (OMC) system for shoulder measurements in a clinical context.
Methods: The validation analyses were conducted on 19 healthy male volunteers. Twelve reflective markers were placed on each participant's trunk, scapula and across the arm and one IMU was attached via a self-adhesive strap on the forearm. A single tester simultaneously collected shoulder kinematic data for four shoulder movements: flexion, extension, external rotation, and abduction. Agreement between OMC system and IMU measurements was assessed with Bland-Altman analyses. Secondary analysis included mean biases, root mean square error (RMSE) analysis and Welch's -test.
Results: Bland-Altman limits of agreement (LoA) exceeded the acceptable range of mean difference for 95% of the population (-22.27°, 11.31°). The mean bias showed high levels of agreement within 8° for all four movements. More than 60% of participants demonstrated mean bias less than 10° between methods. Statistically significant differences were found between measurements for abduction (P<0.001) and flexion (P=0.027) but not for extension and external rotation (P≥0.05).
Conclusions: Our study shows preliminary evidence for acceptable accuracy of a commercially available IMU against an OMC system for assessment of shoulder movements by a single tester. The IMU also exhibits similar whole degree of error compared to a standard goniometer with potential for application in remote rehabilitation.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9634209 | PMC |
http://dx.doi.org/10.21037/mhealth-22-7 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!