Thermal injury from forced-air warmer device precipitated by massive extravasation from peripheral venous cannula.

Saudi J Anaesth

Department of Neurosurgery, King Fahad Medical City, Riyadh, Kingdom of Saudi Arabia.

Published: September 2022

Extravasation injury (EVI) is an iatrogenic complication of venous cannulation. Usually innocuous but occasionally it engenders sequelae. Its severity is determined by various physicochemical properties of infusate. A 50-year-old patient developed leg EVI from crystalloids infused through a pressurized digital infuser (PDI), likely from cannula tip displacement during positioning for craniotomy. We ignored checking gravity-aided free-fluids flow before switching on PDI. Following surgery, the patient had an edematous leg with bullae and epidermal peelings from severe extravasation and burns, respectively. Doppler revealed patent leg arteries. Therefore, EVI was conservatively managed, with complete recovery. Apparently, increased local tissue pressure from extravasation produced conditions of peripheral circulation sufficiency predisposing the leg to thermal injury from the forced-air warmer. On inspecting PDI postoperatively, its upper-pressure alarm limit was 300 mmHg, which prevented it from sounding alarm during extravasation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9630688PMC
http://dx.doi.org/10.4103/sja.sja_786_21DOI Listing

Publication Analysis

Top Keywords

thermal injury
8
injury forced-air
8
forced-air warmer
8
extravasation
5
warmer device
4
device precipitated
4
precipitated massive
4
massive extravasation
4
extravasation peripheral
4
peripheral venous
4

Similar Publications

Background: High-temperature environment can cause acute kidney injury affecting renal filtration function. To study the mechanism of renal injury caused by heat stress through activates TLR4/NF-κB/NLRP3 signaling pathway by disrupting the filtration barrier in broiler chickens. The temperature of broilers in the TN group was maintained at 23 ± 1 °C, and the HS group temperature was maintained at 35 ± 1℃ from the age of 21 days, and the high temperature was 10 h per day, and one broiler from each replicate group at the age of 35 and 42 days was selected for blood sampling, respectively.

View Article and Find Full Text PDF

Mepiquat is a contaminant produced in thermal-processed food. It can induce spleen and liver injury. However, the mechanism that mepiquat induced hepatotoxicity remains unclear.

View Article and Find Full Text PDF

Burn injuries remain a major clinical problem worldwide, which require special management by experienced plastic surgeons. However, they cannot be available in every healthcare unit; consequently, there is a need for effective treatment options that could be utilized by a wide range of non-expert healthcare professionals. The aim of the present experimental study was to investigate the safety and efficacy of using a fibrin sealant (TISSEEL) compared to the conventional treatment with sulfadiazine on partial-thickness burn in a rat animal model.

View Article and Find Full Text PDF

Enhanced Photothermal/Immunotherapy under NIR Irradiation Based on Hollow Mesoporous Responsive Nanomotor.

Inorg Chem

December 2024

Laboratory of Electromechanical Integrated Manufacturing of High-performance Electronic Equipment, School of Mechano-Electronic Engineering, School of Life Science and Technology, Xidian University, Xi'an, Shaanxi 710071, China.

In this research, a hollow mesoporous responsive nanomotor was proposed for enhanced photothermal/immunotherapy under near infrared (NIR) irradiation. HA-HMCuS/AS as the nanomotor composed of hollow mesoporous copper sulfide (HMCuS) loaded with artesunate (AS) and hyaluronic acid (HA) was utilized to induce the polarization of tumor-associated macrophages. At the beginning, ResNet18 deep learning model was utilized to predict the Brunauer-Emmett-Teller (BET) surface area of HMCuS based on the morphology data set which was obtained from our conventional research.

View Article and Find Full Text PDF

This is an annual report indicating the number and early clinical results of annual vascular treatment performed by vascular surgeons in Japan in 2018, as analyzed by database management committee (DBC) members of the Japanese Society for Vascular Surgery (JSVS). To survey the current status of vascular treatments performed by vascular surgeons in Japan, the DBC members of the JSVS analyzed the vascular treatment data provided by the National Clinical Database (NCD), including the number of treatments and early results such as operative and hospital mortality. In total, 143,745 vascular treatments were registered by 1,090 institutions in 2018.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!