Background: Occlusal force represents masticatory function. Using quantifiable occlusal indicators provides a more objective occlusal force evaluation. In the recent dental practice, digital methods such as the Dental Prescale II (DP2, GC Corp., Tokyo, Japan) and T-scan (T-Scan III v8; Tekscan Inc.) are commonly used in clinics to evaluate treatment outcomes. The T-scan provides the relative bite force (%) compared to the maximal bite force on individual teeth or the unilateral arch. The DP2 can quantify occlusal force, measured in newtons (N), on the half arch or the overall bite, but it is difficult to identify the bite force on an individual tooth. It is difficult to select a device that fulfils all the requirements to record occlusal force. This study aimed to investigate the association between the bite measured by the DPS2 and T-scan to determine whether the measured bite force is comparable through calculation.

Methods: A total of 80 healthy adults, including 41 women and 39 men with a mean age of 38.2, were requested to bite pressure sensitive film sheets ten minutes apart. Linear regression analysis was used to estimate the measured bite force by the DP2 and T-scan.

Results: There was a significant positive correlation between the occlusal force measured by the DP2 and T-scan (P < 0.01) when intercept was equal to zero as confounders were adjused. These results provided the comparability of the measured occlusal forces determined by the DP2 and T-scan.

Conclusion: The estimated bite force determined by DP2 and T-Scan is convertible using the linear equation from this study to increase the value for clinical applications. The estimated bite force from the two quantifiable occlusal indicators are comparable. The two commercially available quantifiable occlusal indicators can be fully adapted to all clinical requirements according to this result.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9636614PMC
http://dx.doi.org/10.1186/s12903-022-02484-9DOI Listing

Publication Analysis

Top Keywords

bite force
32
occlusal force
20
force
13
quantifiable occlusal
12
occlusal indicators
12
bite
11
occlusal
9
correlation occlusal
8
force individual
8
force measured
8

Similar Publications

Low-level laser therapy (LLLT) is known for its biostimulant properties, which can reduce inflammation and promote tissue regeneration. The present study is randomized, blinded, and placebo-controlled and aims to investigate the role of LLLT in the postoperative recovery of facial fractures. Patients with fractures of the zygomatic bone are selected and divided into two groups: low-level laser and red placebo light.

View Article and Find Full Text PDF

Objective: To study the biomechanical changes induced by differences in perioral force in patients with missing anterior maxillary teeth at rest via finite element analysis (FEA).

Methods: Using conical beam CT (CBCT) images of a healthy person, models of the complete maxillary anterior dental region (Model A) and maxillary anterior dental region with a missing left maxillary central incisor (Model B) were constructed. The labial and palatine alveolar bone and tooth surface of the bilateral incisor and cusp regions were selected as the application sites, the resting perioral force was applied perpendicular to the tissue surface, and the changes in maxillary stress and displacement after the perioral force was simulated were analyzed.

View Article and Find Full Text PDF

Background: Immature maxillary central teeth can be managed by using several treatment options. The aim of this finite element stress analysis study was to evaluate the effect of different treatment procedures on the stresses on immature maxillary incisor teeth models that generated on cone beam computed tomography, by trauma and bite forces.

Methods: A total of 11 different models consisting of revascularization treatment using MTA and biodentine and the state of the root apex formed with cement after treatment, apexification, modified apexification, traditional root canal treatment and two different control groups have been created.

View Article and Find Full Text PDF

The forces generated by action potentials in muscle cells shuttle blood, food and waste products throughout the luminal structures of the body. Although non-invasive electrophysiological techniques exist, most mechanosensors cannot access luminal structures non-invasively. Here we introduce non-toxic ingestible mechanosensors to enable the quantitative study of luminal forces and apply them to study feeding in living Caenorhabditis elegans roundworms.

View Article and Find Full Text PDF

Background: Rupture of the lumbar catheter in lumboperitoneal (LP) shunts is rare and typically occurs due to long-term mechanical stress. The authors describe an unusual case of early postoperative lumbar catheter severing after a fall on the buttocks.

Observations: A 78-year-old woman underwent LP shunt placement for communicating hydrocephalus after aneurysmal subarachnoid hemorrhage.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!