Background: Atrial fibrillation and flutter are often treated with external electrical cardioversion (ECV) in patients with potentially electrically sensitive cardiovascular implantable electronic devices (CIED). Long-term follow-up data on contemporary CIED undergoing ECV is sparse. The aim is to investigate shock-related complications and impact on CIEDs.

Methods: All ECV procedures from 2010 to 2020 in patients with CIED performed at a tertiary university hospital were identified in the Danish National Patient Registry. Changes in device measurements after ECV were retrospectively studied and procedure-related complications were identified by review of medical records.

Results: We analyzed 763 ECV procedures in 372 patients, median device implant time 1.9 years. The mean age of patients was 69.9 ± 9.9 years of which 73.4% were men. We identified two cases of device programming changes and four cases of premature battery depletion (≤3 years after device implant). Minor changes in device measurements were found for impedances, sensing, and pacing thresholds. No patients died due to ECV-related device dysfunctions within the first 12 months after cardioversions.

Conclusion: External cardioversion in patients with contemporary pacemakers and implantable cardioverter-defibrillators seems safe in the majority of patients. Clinically important changes in device function following cardioversion were rarely observed but may be critical for device function. In an observational study, causality between cardioversion and device dysfunction cannot be established. For patient safety, we suggest that routine device interrogation after cardioversion still should be part of standard care.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pace.14616DOI Listing

Publication Analysis

Top Keywords

changes device
12
device
10
external cardioversion
8
atrial fibrillation
8
fibrillation flutter
8
patients
8
ecv procedures
8
device measurements
8
device implant
8
device function
8

Similar Publications

Purpose: Atrial fibrillation (AF) is the most common chronic cardiac arrhythmia that increases the risk of stroke, primarily due to thrombus formation in the left atrial appendage (LAA). Left atrial appendage occlusion (LAAO) devices offer an alternative to oral anticoagulation for stroke prevention. However, the complex and variable anatomy of the LAA presents significant challenges to device design and deployment.

View Article and Find Full Text PDF

Introduction: Altered vascular microcirculation is recognized as a risk factor for anastomotic leakage (AL) in colorectal surgery. However, few studies evaluated its impact on AL using different devices, with heterogeneous results. The present study reported the initial experience measuring gut microcirculatory density and flow with the aid of incidence dark-field (IDF) videomicroscopy (Cytocam, Braedius, Amsterdam, The Netherlands) comparing its operative outcome using a propensity score matching (PSM) model based on age, gender, and Charlson Comorbidity Index (CCI).

View Article and Find Full Text PDF

Objective: Patellofemoral pain syndrome (PFPS) is a common knee issue, and hip joint function significantly affects knee health. Gluteus activation exercises are a promising treatment for PFPS. This study aims to investigate the impact of gluteal muscle activation exercises on the muscle involvement and movement patterns of young male patients with PFPS.

View Article and Find Full Text PDF

Background:  Digital health interventions targeting behavior change are promising in adults and adolescents; however, less attention has been given to younger children. The proliferation of wearables, such as smartwatches and activity trackers, that support the collection of and reflection on personal health data highlights an opportunity to consider novel approaches to supporting health in young children (aged 5-11 y).

Objective:  This review aims to investigate how smartwatches and activity trackers have been used across child health interventions (for children aged 5-11 y) for different health areas, specifically to identify the population characteristics of those being targeted, describe the characteristics of the devices being used, and report the feasibility and acceptability of these devices for health-related applications with children.

View Article and Find Full Text PDF

Disturbance in sleep and activity rhythms are significant health risks associated with alcohol use during adolescence. Many investigators support the theory of a reciprocal relationship between disrupted circadian rhythms, sleep patterns, and alcohol usage. However, in human studies it is difficult to disentangle other factors (i.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!