Introduction: Up to one-third of patients indicated for transcatheter aortic valve implantation (TAVI) may be unsuitable for transfemoral TAVI (TF-TAVI) according to manufacturers' recommendations and numerous professional societies.
Objective: This study aimed to investigate the predictive value of manufacturers' guidelines for major vascular access site complications using the Perclose ProGlide device.
Methods: Among 208 patients undergoing TF-TAVI, 144 patients (69.2%) were deemed eligible for TF-TAVI according to the manufacturer's instructions. A minimal lumen diameter (MLD) of the femoral artery below the manufacturer's specified limits and/or the presence of circumferential calcification were the reasons for ineligibility. Calcium score (CS), sheath-to-femoral artery ratio (SFAR) and MLD were estimated from computed tomography imaging. Vascular complications (VCs) (defined according to VARC-2 criteria) were retrospectively compared.
Results: Patients in the ineligible group had higher SFAR (1.13±0.15 vs. 0.88±0.107, p<0.001) and CS (1.66±0.99 vs. 1.24±0.73; p=0.003), and significantly lower MLD (7.72±1.03 vs. 6.31±0.96 mm; p<0.001) compared to the eligible group. Major (6.3% vs. 12.3%, p=0.13) and minor VCs (10.4% vs. 15.6%, p=0.29) were similar in the eligible and ineligible groups. The ineligible group had higher rates of rupture (0.7% vs. 6.3%; p=0.03). SFAR was the only independent predictor of major VCs (OR 469.1, 95% CI 4.95-44466.57, p=0.008).
Conclusion: The TAVI team should not decide whether the patient is suitable for a femoral approach based solely on the manufacturer's criteria, and should incorporate additional factors that could be predictive of major VCs.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.repc.2021.09.019 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!