Grounded in hydrodynamic theory, proximal isovelocity surface area (PISA) is a simplistic and practical technique widely used to quantify valvular regurgitation flow. PISA provides a relatively reasonable, though slightly underestimated flow rate for circular orifices. However, for elliptical orifices frequently seen in functional mitral regurgitation, PISA underestimates the flow rate. Based on data obtained with computational fluid dynamics (CFD) and in vitro experiments using systematically varied orifice parameters, we hypothesized that flow rate underestimation for elliptical orifices by PISA is predictable and within a clinically acceptable range. We performed 45 CFD simulations with varying orifice areas 0.1, 0.3 and 0.5 cm, orifice aspect ratios 1:1, 2:1, 3:1, 5:1, and 10:1, and peak velocities (V) 400, 500 and 600 cm/s. The ratio of computed effective regurgitant orifice area to true effective area (EROA/EROA) against the ratio of aliasing velocity to peak velocity (V/V) was analyzed for orifice shape impact. Validation was conducted with in vitro imaging in round and 3:1 elliptical orifices. Plotting EROA/EROA against V/V revealed marginal flow underestimation with 2:1 and 3:1 elliptical axis ratios against a circular orifice (< 10% for 8% V/V), rising to ≤ 35% for 10:1 ratio. In vitro modeling confirmed CFD findings; there was a 8.3% elliptical EROA underestimation compared to the circular orifice estimate. PISA quantification for regurgitant flow through elliptical orifices produces predictable, but generally small, underestimation deemed clinically acceptable for most regurgitant orifices.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s10554-022-02729-2 | DOI Listing |
Heliyon
January 2025
Institute of Mathematics, Henan Academy of Sciences, Zhengzhou, 450046, China.
This study examines the behavior of the Casson nanofluid bioconvection flow around a spinning disc under various influences, including gyrotactic microorganisms, multiple slips, and thermal radiation. Notably, it accounts for the reversible nature of the flow and incorporates the esterification process. The aim of this study is to investigate the influence of reversible chemical reactions on the flow behavior of a Casson nanofluid in the presence of bioconvective microorganisms over a spinning disc.
View Article and Find Full Text PDFInt J Exerc Sci
December 2024
Exercise Science, Florida Southern College, Lakeland, FL, USA.
The purpose of the study was to compare heart rate (HR), systolic blood pressure (SBP), and diastolic blood pressure (DBP) following high load resistance exercise (HLRE) and blood flow restriction exercise (BFRE) with a knee wrap (kBFRE) and pneumatic cuff (pBFRE). Eleven men (N = 9) and women (N = 2) participated. HR, SBP, and DBP were collected at Rest, immediately post exercise (IP), 10-, 30-, and 45-minutes post exercise.
View Article and Find Full Text PDFJ Soc Cardiovasc Angiogr Interv
December 2024
The University of Toledo, Toledo, Ohio.
Background: Atherosclerotic renal artery stenosis (ARAS) may provoke hypertension and/or impaired kidney function. Some patients develop uncontrolled hypertension and deteriorating kidney function despite optimal medical therapy. In these patients, endovascular treatment is an important therapeutic option.
View Article and Find Full Text PDFAnticancer Agents Med Chem
January 2025
Department of Biology, Faculty of Science, Atatürk University, Erzurum, Turkey.
Introduction/objective: Several nutraceuticals, food, and cosmetic products can be developed using royal jelly. It is known for its potential health benefits, including its ability to boost the immune system and reduce inflammation. It is rich in vitamins, minerals, and antioxidants, which can improve general health.
View Article and Find Full Text PDFJ Sep Sci
January 2025
R & D Laboratory, TASNEE Technology and Innovation Centre, Al-Jubail Industrial City, Saudi Arabia.
In this study, a commercially available polypropylene homopolymer (H-PP) was blended with blow molding polyethylene (PE) grade via melt mixing using a compounding machine. The resulting blends were subjected to high-temperature size exclusion chromatography (SEC) analysis, coupled with infrared-5 (IR-5), viscometer (VISCO), and multi-angle laser light scattering (MALS) detectors. The molecular weight (MW) and MW distributions were investigated using SEC, and the exact blend compositions were evaluated using C nuclear magnetic resonance.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!