Objectives: In this study, we characterise the imaging-mode performance of the positron emission tomography (PET) subsystem of the RefleXion X1 machine using the NEMA NU-2 2018 standard.
Methods: The X1 machine consists of two symmetrically opposing 90 arcs of PET detectors incorporated into the architecture of a ring-gantry linear accelerator rotating up to 60 RPM. PET emissions from a tumour are detected by the PET detectors and used to guide the delivery of radiation beam. Imaging performance of the PET subsystem on X1 machine was evaluated based on sensitivity of the PET detectors, spatial resolution, count-loss performance, image quality, and daily system performance check.
Results: PET subsystem sensitivity was measured as 0.183 and 0.161 cps/kBq at the center and off-center positions, respectively. Spatial resolution: average FWHM values of 4.3, 5.1, and 6.7 mm for the point sources at 1, 10, and 20 cm off center, respectively were recorded. For count loss, max NECR: 2.63 kcps, max true coincidence rate: 5.56 kcps, and scatter fraction: 39.8%. The 10 mm sphere was not visible. Image-quality contrast values were: 29.6%, 64.9%, 66.5%, 81.8%, 81.2%, and background variability: 14.8%, 12.4%, 10.3%, 8.8%, 8.3%, for the 13, 17, 22, 28, 37 mm sphere sizes, respectively.
Conclusions: When operating in an imaging mode, the spatial resolution and image contrast of the X1 PET subsystem were comparable to those of typical diagnostic imaging systems for large spheres, while the sensitivity and count rate were lower due to the significantly smaller PET detector area in the X1 system. Clinical efficacy when used in BgRT remains to be validated.
Advances In Knowledge: This is the first performance evaluation of the PET subsystem on the novel BgRT machine. The dual arcs rotating PET subsystem on RefleXion X1 machine performance is comparable to those of the typical diagnostic PET system based on the spatial resolution and image contrast for larger spheres.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10997023 | PMC |
http://dx.doi.org/10.1259/bjr.20220387 | DOI Listing |
Nucl Instrum Methods Phys Res A
June 2024
Department of Radiology, University of Chicago, Chicago, IL 60637, USA.
We report the development of a high-sensitivity and high-resolution PET subsystem for a next-generation preclinical PET/EPR hybrid scanner for investigating and improving hypoxia imaging with PET. The PET subsystem consists of 14 detector modules (DM) installed within a cylindrical supporting frame whose outer and inner diameters are 115mm and 60mm, respectively. Each DM contains eight detector units (DU) in a row and each DU is made of a 12×12 array of 1×1×10mm LYSO crystals (with a 1.
View Article and Find Full Text PDFBrain Commun
January 2024
Department of Radiology, Mayo Clinic, Rochester, MN 55905, USA.
Disruption of the default mode network is a hallmark of Alzheimer's disease, which has not been extensively examined in atypical phenotypes. We investigated cross-sectional and 1-year longitudinal changes in default mode network sub-systems in the visual and language variants of Alzheimer's disease, in relation to age and tau. Sixty-one amyloid-positive Alzheimer's disease participants diagnosed with posterior cortical atrophy ( = 33) or logopenic progressive aphasia ( = 28) underwent structural MRI, resting-state functional MRI and [F]flortaucipir PET.
View Article and Find Full Text PDFMed Phys
April 2024
Medical Physics Program, University of Nevada, Las Vegas, Nevada, USA.
Background: In spite of the tremendous potential of game-changing biological image- and/or biologically guided radiation therapy (RT) and adaptive radiation therapy for cancer treatment, existing limited strategies for integrating molecular imaging and/or biological information with RT have impeded the translation of preclinical research findings to clinical applications. Additionally, there is an urgent need for a highly integrated small-animal radiation therapy (SART) platform that can seamlessly combine therapeutic and diagnostic capabilities to comprehensively enhance RT for cancer treatment.
Purpose: We investigated a highly integrated quad-modal on-board imaging configuration combining positron emission tomography (PET), single-photon emission computed tomography (SPECT), photon-counting spectral CT, and cone-beam computed tomography (CBCT) in a SART platform using a Monte Carlo model as a proof-of-concept.
Neurobiol Dis
December 2023
Key Laboratory of Neuropharmacology and Translational Medicine of Zhejiang Province, School of Pharmaceutical Sciences, Zhejiang Chinese Medical University, Hangzhou 310053, China; Zhejiang Rehabilitation Medical Center, The Third Affiliated Hospital of Zhejiang Chinese Medical University, Hangzhou 310061, China. Electronic address:
The raphe nuclei, the primary resource of forebrain 5-HT, play an important but heterogeneous role in regulating subcortical excitabilities. Fundamental circuit organizations of different median raphe (MR) subsystems are far from completely understood. In the present study, using cell-specific viral tracing, Ca fiber photometry and epilepsy model, we map out the forebrain efferent and afferent of different MR Pet subpopulations and their divergent roles in epilepsy.
View Article and Find Full Text PDFMed Phys
April 2024
Department of Engineering Physics, Tsinghua University, Beijing, China.
Background: A compact PET/SPECT/CT system Inliview-3000B has been developed to provide multi-modality information on small animals for biomedical research. Its PET subsystem employed a dual-layer-offset detector design for depth-of-interaction capability and higher detection efficiency, but the irregular design caused some difficulties in calculating the normalization factors and the sensitivity map. Besides, the relatively larger (2 mm) crystal cross-section size also posed a challenge to high-resolution image reconstruction.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!