Purpose: To assess the reliability of successive Corvis ST® measurements (CST, Oculus, Wetzlar, Germany) in keratoconus (KC) ≥ 2 years after accelerated corneal crosslinking (9 mW/cm, 10 min, 5.4 J/cm) compared to untreated KC corneas.

Methods: Three successive CST measurements per eye were performed in ≥ 2 years after CXL (CXLG, n = 20 corneas of 16 patients) and a control group consisting of non-operated, ABC-stage-matched KC corneas according to Belin's ABCD KC grading (controls, n = 20 corneas, 20 patients). Main outcome measures included maximal keratometry (Kmax), the Belin/Ambrósio-Enhanced-Ectasia-Deviation-Index BAD-D; the biomechanical parameters A1 velocity, deformation amplitude (DA) ratio 2 mm, Ambrósio relational thickness to the horizontal profile (ARTh), integrated radius, stiffness parameter A1 (SP-A1), and the Corvis Biomechanical Factor (CBiF, the linearized term of the Corvis Biomechanical Index). Mean values, standard deviations, and Cronbach's alpha (CA) were calculated.

Results: Both groups were tomographically comparable (BAD: 11.5 ± 4.7|11.2 ± 3.6, p = 0.682, Kmax: 60.5 ± 7.2|60.7 ± 7.7, p = 0.868 for controls|CXLG, paired t-test). A1 velocity (mean ± SD: 0.176 ± 0.02|0.183 ± 0.02, p = 0.090, CA: 0.960|0.960), DA ratio 2 mm (6.04 ± 1.13|6.14 ± 1.03, p = 0.490, CA: 0.967|0.967), integrated radius (12.08 ± 2.5|12.42 ± 1.9, p = 0.450, CA: 0.976|0.976), and CBiF (4.62 ± 0.6|4.62 ± 0.4, p = 0.830, CA: 0.965|0.965) were also comparable (controls|CXLG). ARTh was significantly higher in controls (177.1 ± 59, CA: 0.993) than after CXL (155.21 ± 65, p = 0.0062, CA: 0.993) and SP-A1 was significantly higher after CXL (59.2 ± 13, CA: 0.912) than in controls (52.2 ± 16, p = 0.0018, CA: 0.912).

Conclusion: ARTh and SP-A1 differed significantly between controls and CXLG. Biomechanical measurements were generally of excellent reliability in both groups. CXL seems to affect biomechanical measurements of human corneas over more than 2 years.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10049946PMC
http://dx.doi.org/10.1007/s00417-022-05881-6DOI Listing

Publication Analysis

Top Keywords

successive corvis
8
corvis st®
8
st® measurements
8
accelerated corneal
8
corneal crosslinking
8
compared untreated
8
n = 20 corneas
8
corneas patients
8
ratio 2 mm
8
integrated radius
8

Similar Publications

Purpose: To assess the reliability of successive Corvis ST® measurements (CST, Oculus, Wetzlar, Germany) in keratoconus (KC) ≥ 2 years after accelerated corneal crosslinking (9 mW/cm, 10 min, 5.4 J/cm) compared to untreated KC corneas.

Methods: Three successive CST measurements per eye were performed in ≥ 2 years after CXL (CXLG, n = 20 corneas of 16 patients) and a control group consisting of non-operated, ABC-stage-matched KC corneas according to Belin's ABCD KC grading (controls, n = 20 corneas, 20 patients).

View Article and Find Full Text PDF

Background: This study assesses the reliability of successive corneal biomechanical response measurements by the Corneal Visualization Scheimpflug Technology (CST, Corvis ST , Oculus Optikgeräte, Wetzlar, Germany) in different keratoconus (KC) stages.

Methods: A total of 173 eyes (15 controls: 15 eyes, and 112 KC patients: stages 1|1-2|2|2-3|3|3-4|4, n = 26|16|36|18|31|26|5 according to Topographical KC Classification, TKC) were repeatedly examined five times with the CST, each after repositioning the patient's head and re-adjusting the device. Tomographical analysis (Pentacam HR ; Oculus, Wetzlar, Germany) was performed once before and once after CST measurements.

View Article and Find Full Text PDF

Purpose: To investigate the characteristics of corneal biomechanics in Chinese preschool children with different refractive status.

Methods: Study participants were 108 Chinese children (216 eyes) aged 4 to 6 years with a spherical equivalent refraction between -9.00 and +9.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!