Trifecta valve is made from single bovine pericardial sheet, which is externally mounted on a titanium stent. This valve design provides good hemodynamic performance and wide effective orifice area. However, there have been some reports of early structural valve dysfunction which caused acute heart failure. Case 1:An 80-year-old man who had undergone aortic valve replacement (AVR) using Trifecta 21 mm five years ago. He presented with fever and diastolic murmur. Echocardiography showed prosthetic valve regurgitation, and the patient later developed prosthetic valve endocarditis. Case 2:An 80-year-old woman who had undergone AVR using Trifecta GT 21 mm two years ago developed sudden chest pain and dyspnea, prosthetic valve regurgitation. Both patients underwent re-AVR with a new bioprosthetic valve via re-sternotomy. Large tear was found in the right coronary cusp of the Trifecta in both cases. During Trifecta accomodation, we should avoid oversizing, and check the position of coronary ostia and sinus of Valsalva. Secure late outcome of Trifecta GT remains unproven and its use still needs caution.
Download full-text PDF |
Source |
---|
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!