Introduction: Early diagnosis and appropriate infection control are important to prevent the transmission of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). In this study, we aimed to assess the diagnostic performance of SARS-CoV-2 rapid antigen detection (RAD) tests and the factors that cause nonspecific reactions.

Methods: Nasopharyngeal swab specimens (n = 100), sputum specimens (n = 10), and lithium-heparin plasma samples (n = 100) were collected. We evaluated Espline®SARS-CoV-2 (Espline) and SARS-CoV-2 Rapid Antigen Test that also known as STANDARD Q® (STANDARD Q), with reverse transcription-polymerase chain reaction (RT-PCR) and Lumipulse® Presto SARS-CoV-2 Ag as reference tests. In addition, we investigated the effects of inadequate pretreatment methods and five potential causes of nonspecific reactions.

Results: The sensitivities of Espline and STANDARD Q were 60% and 57%, respectively, and their specificity was 100%. It was confirmed that the judgment line for the positive insufficiently mixed specimens was faint. A false-positive result was observed with STANDARD Q when sputum was used as a specimen to investigate judgment the effect of viscosity.

Conclusions: Espline and STANDARD Q show good sensitivity for specimens with Ct values less than 25, but specimens collected within 9 days of symptom onset may still give false negatives. The test should be performed carefully, and the results should be judged comprehensively, taking into account clinical symptoms and patient background.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9595385PMC
http://dx.doi.org/10.1016/j.jiac.2022.10.007DOI Listing

Publication Analysis

Top Keywords

sars-cov-2 rapid
12
rapid antigen
12
diagnostic performance
8
antigen detection
8
espline standard
8
sars-cov-2
5
specimens
5
standard
5
analysis diagnostic
4
performance factors
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!