Background: Accurate measurement of glycated hemoglobin (HbA1c) is crucial for a diabetes diagnosis and subsequent patient management. The detection method and presence of variant Hb can interfere with HbA1c measurements. We evaluated the HbA1c-measuring performance of the DxC 700 AU (Beckman Coulter, Brea, CA, USA) immunoassay-based device in comparison with another immunoassay device and the reference method.
Methods: A total of 120 normal and 14 variant Hb samples were analyzed using the Cobas c 513 (Roche Diagnostics, Mannheim, Germany) and DxC 700 AU analyzers. Variant Hb samples were also analyzed using the reference method, along with 20 normal samples. The accuracy, precision, linearity, and carryover were determined.
Results: DxC 700 AU results strongly correlated with those of Cobas c 513 and exhibited accuracy in comparison with the reference method. The within-run, between-run, between-day, and total imprecision (%CV) values for the low- and high-concentration control materials were below 2%. The results of DxC 700 AU were linear over a wide HbA1c range (3.39%-18.30%). Although DxC 700 AU performed well in the presence of variant Hb, the HbA1c concentration was underestimated in the presence of fetal Hb. The possibility of interference from a high HbH proportion could not be ruled out.
Conclusions: The overall analytical performance of DxC 700 AU was acceptable. The device is accurate, precise, and linear over a wide HbA1c concentration range. Although DxC 700 AU results highly correlated with those of Cobas c 513, caution should be exercised in cases of high HbF and HbH concentrations.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9618898 | PMC |
http://dx.doi.org/10.3343/alm.2023.43.2.167 | DOI Listing |
Sci Rep
June 2024
Department of Clinical Chemistry, School of Biomedical and Laboratory Sciences, College of Medicine and Health Sciences, University of Gondar, P.O. Box 196, Gondar, Ethiopia.
Metabolic syndrome (MetS) poses a significant public health challenge globally, including in Ethiopia, with risks for both mothers and children. Unfortunately, there is limited data on MetS in pregnant Ethiopian women. This study aims to evaluate the prevalence and factors associated with MetS in this population.
View Article and Find Full Text PDFJ Appl Lab Med
May 2024
Clinical Laboratory, Bassett Medical Center, Cooperstown, NY, United States.
Background: Hemolysis is a common reason for specimen rejection in the laboratory. Our experience suggested that hemolysis (H) flag limits are too strict for some analytes leading to unnecessary specimen rejections. This study summarizes H flags for commonly rejected analytes on the Beckman Coulter DxC 700 AU analyzer.
View Article and Find Full Text PDFPLoS One
August 2023
Department of Clinical Chemistry, School of Biomedical and Laboratory Sciences, College of Medicine and Health Sciences, University of Gondar, Gondar, Ethiopia.
Introduction: Hormonal contraceptives are artificial preparations that contain artificial progestins and Ethinylestradiol; these preparations are utilized by women of reproductive age to prevent pregnancy. Roughly a billion women in the world use some form of contraceptive worldwide. Despite the utility of these preparations, they are linked with several adverse effects, including disturbances of liver functionality and integrity.
View Article and Find Full Text PDFAnn Lab Med
March 2023
Department of Laboratory Medicine, Yonsei University College of Medicine, Seoul, Korea.
Background: Accurate measurement of glycated hemoglobin (HbA1c) is crucial for a diabetes diagnosis and subsequent patient management. The detection method and presence of variant Hb can interfere with HbA1c measurements. We evaluated the HbA1c-measuring performance of the DxC 700 AU (Beckman Coulter, Brea, CA, USA) immunoassay-based device in comparison with another immunoassay device and the reference method.
View Article and Find Full Text PDFJ Multidiscip Healthc
July 2022
Department of Clinical Chemistry, School of Biomedical and Laboratory Sciences, College of Medicine and Health Sciences, University of Gondar, Gondar, Ethiopia.
Background: Dyslipidemia refers to a lipid profile disturbance due to decreased high-density lipoprotein cholesterol and elevated low-density lipoprotein cholesterol, triglycerides, and total cholesterol. infection can lead to some appetite-related disorders that may cause deregulated absorption of nutrients in the digestive system, contributing to changes in serum lipids. The purpose of this study is to assess dyslipidemia and its associated factors among -infected patients attending at University of Gondar Comprehensive Specialized Hospital.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!