Background: Ablation of atrial tachyarrhythmia in adults with congenital heart disease (ACHD) is challenging because of complex anatomy and high scar burden. We proposed that the addition of high-density mapping with the PentaRay® (Biosense Webster, Inc) mapping catheter (EAM+P) to 3-dimensional electroanatomic mapping (EAM) allows for rapid acquisition of high-resolution maps and shorter procedure times.
Methods: In this single-center, retrospective cohort study of patients with ACHD who underwent atrial arrhythmia ablation, patients were divided those who underwent ablation with EAM and those who underwent ablation with EAM+P.
Results: Fifteen ablations were performed in 13 patients using standard EAM, and 11 ablations were performed in 10 patients using EAM+P. There was no difference in mean age or complexity of congenital heart disease. The procedure duration was 1.5 times longer in the EAM than in the EAM+P group (P = .015). The dose area product was 12 times higher in the EAM than in the EAM+P group (P = .001). A higher number of venous access sites were used for EAM cases than for EAM+P cases (P = .008). Acute success rates of ablation and recurrence rates at 1 year were similar in the 2 groups. There were no procedure-related complications in either group.
Conclusion: This is the first study to evaluate the use of the PentaRay® high-density mapping catheter for ablation of atrial tachyarrhythmia in patients with ACHD. The use of the PentaRay® high-density mapping catheter results in shorter procedure time, decreased radiation exposure, and fewer venous access sites.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9632387 | PMC |
http://dx.doi.org/10.14503/THIJ-20-7535 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!