Impact of manual crimping on stapedotomy outcomes.

J Laryngol Otol

Department of Otorhinolaryngology, Head and Neck Surgery, Luzerner Kantonsspital, Lucerne, Switzerland.

Published: September 2023

Background: The impact of tight stapes crimping on hearing is a matter of debate. Several studies postulate that tight crimping is essential for lifelong success, whereas others have debated whether firm attachment leads to incus necrosis. Several types of prostheses with different coupling mechanisms have been developed, and manual crimping remains the most frequently used technique. This study investigates whether tightness really does affect hearing outcome.

Methods: The hearing results of patients who underwent primary stapedotomies using three different titanium pistons were analysed. The surgeons categorised the firmness of the piston attachment into 'tight' and 'loose' crimping groups. Hearing outcome and reasons for revision surgical procedures were investigated.

Results: The mean post-operative air-bone gap for frequencies of 0.5-4 kHz was 8.80 dB for the tight crimping group ( = 308) and 9.55 dB for the loose crimping group ( = 39). No significant difference was found ( = 0.4650). Findings at revision procedures were comparable (1.6 per cent 5 per cent).

Conclusion: Although firm crimping is strongly advised, a movable loop upon palpation does not lead to unsatisfactory hearing results, and does not mandate piston replacement or bone cement use.

Download full-text PDF

Source
http://dx.doi.org/10.1017/S0022215122002316DOI Listing

Publication Analysis

Top Keywords

crimping
8
manual crimping
8
tight crimping
8
crimping group
8
hearing
5
impact manual
4
crimping stapedotomy
4
stapedotomy outcomes
4
outcomes background
4
background impact
4

Similar Publications

This study investigates the impact of carding and blending recycled carbon fibers (rCF) with crimped thermoplastic polypropylene (PP) fibers on the mechanical properties of rCF, using a Weibull statistical approach. Tensile properties of rCF were evaluated before and after carding with varying rCF/PP blend ratios (100/0%, 85/15%, 70/30%, and 50/50%). A comparison between the two-parameter and three-parameter Weibull models showed that the two-parameter model provided a better fit for rCF properties before carding.

View Article and Find Full Text PDF

The integration of self-expandable nitinol frames with cable-driven parallel mechanisms offers a promising advancement in minimally invasive cardiovascular interventions. This study presents the design, fabrication, and verification of a miniaturized self-expandable nitinol frame to enhance catheter tip steerability and navigation within complex vascular anatomies. The frame is reduced in size for delivery through 7-8 Fr sheaths while accommodating diverse vascular diameters, allowing up to a maximum expansion of 15 mm.

View Article and Find Full Text PDF

Malignant biliary obstruction presents a significant therapeutic challenge and has serious consequences including cholangitis and death. Clinically, biliary stenting using self-expanding metallic- stent(SEMS) relieves this obstruction. However, stent occlusion occurs with time due to tumor/epithelial in-growth and bacterial colonization.

View Article and Find Full Text PDF

Aortic valve leaflet assessment to inform novel bioinspired materials: Understanding the impact of collagen fibres on the tissue's mechanical behaviour.

J Mech Behav Biomed Mater

December 2024

Trinity Centre for Biomedical Engineering, Trinity Biomedical Sciences Institute, Trinity College Dublin, 2, Dublin, Ireland; Discipline of Mechanical, Manufacturing, and Biomedical Engineering, School of Engineering, Trinity College Dublin, 2, Dublin, Ireland; Advanced Materials and Bioengineering Research Centre (AMBER), Trinity College Dublin, Ireland. Electronic address:

Aortic stenosis is a prevalent disease that is treated with either mechanical or bioprosthetic valve replacement devices. However, these implants can experience problems with either functionality in the case of mechanical valves or long-term durability in the case of bioprosthetic valves. To enhance next generation prosthetic valves, such as biomimetic polymeric valves, an improved understanding of the native aortic valve leaflet structure and mechanical response is required to provide much needed benchmarks for future device development.

View Article and Find Full Text PDF

The article provides information about strengthening cold-formed thin-walled steel beams made of the sigma profile. An innovative concept for sectional transverse strengthening of thin-walled beams subjected to concentrated forces was investigated. The proposed solution's novelty lies in attaching the sectional transverse strengthening to the beam's cross-section, employing a point crimping technique.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!