Uncertainty of measurement (UM) provides a quantitative estimate for traceability of test results. The Nordtest guide was applied for calculating UM of 26 analytes. For this, internal and external quality control data from July 2019 to April 2020 was used. UM of test results were compared to %TEa values of CLIA '2019, RiliBÄK, and Ricos. It was observed that UM for all analytes were below %TEa values of RiliBÄK. UM value of Albumin, Calcium and Sodium could not meet CLIA '2019 and Ricos guidelines. For results of Albumin, Calcium and Sodium to be traceable, more frequent quality control protocols resulted in decrease in bias. Quality goals were set for these three parameters. This helped in reduction of quality control cycles and optimum utilization of resources.