A large range of laser-matter applications employ ultrashort pulses and high laser intensity. Such processes can lead to unrequired X-ray generation, which represents a hazardous radiation factor even for common laboratory research-grade laser systems. We present here an analysis of the radiation dose rate and X-ray spectrum emitted during ablation of a rotating copper cylinder with respect to several laser parameters. The results show that focused sub-picosecond pulses with intensity above 10 W/cm can exceed the annual irradiation limit even in one hour, requiring appropriate shielding for the safety of the researchers.

Download full-text PDF

Source
http://dx.doi.org/10.1364/OE.468135DOI Listing

Publication Analysis

Top Keywords

research-grade laser
8
laser systems
8
potential hazards
4
hazards mitigation
4
mitigation x-ray
4
x-ray radiation
4
radiation generated
4
generated laser-induced
4
laser-induced plasma
4
plasma research-grade
4

Similar Publications