Cortisol is a steroid hormone involved in a wide range of medical conditions. The level of the hormone fluctuates over time, but with traditional laboratory-based assays, such dynamics cannot be monitored in real time. Here, a reversible cortisol sensor is reported that allows continuous monitoring of cortisol in blood plasma using sampling by microdialysis. The sensor is based on measuring single-molecule binding and unbinding events of tethered particles. The particles are functionalized with antibodies and the substrate with cortisol-analogues, causing binding and unbinding events to occur between particles and substrate. The frequency of binding events is reduced when cortisol is present in the solution as it blocks the binding sites of the antibodies. The sensor responds to cortisol in the high nanomolar to low micromolar range and can monitor cortisol concentrations over multiple hours. Results are shown for cortisol monitoring in filtered and in microdialysis-sampled human blood plasma.