AI Article Synopsis

  • The study examined how different surgical instruments and techniques affected the durability of ventriculoperitoneal shunt catheters, focusing on their ability to withstand assembly and disassembly.
  • Researchers conducted failure testing on a total of 160 catheters, comparing two techniques (pushing vs. pulling) and assessing various instruments to determine their impact on catheter failure rates.
  • Results showed that using certain instruments, like mosquito forceps, significantly reduced failure rates, and the pushing technique resulted in a much lower failure rate compared to the pulling technique (5% vs 81%).*

Article Abstract

Objective: Current surgical techniques use common surgical instruments for sterile shunt assembly. This study investigated the impact of using these techniques and surgical instruments on the mechanical integrity of the ventriculoperitoneal shunt system, specifically shunt catheters.

Methods: The authors conducted failure testing on 85 rifampin-coated catheters and 85 barium-impregnated catheters using 5 different surgical instruments and 2 different surgical techniques. In technique A, the distal end of the catheter was pushed onto the shunt valve inlet connector with the surgical instrument. In technique B, the catheter was pulled over the inlet connector. One hundred sixty catheters underwent 10-repetition-maximum testing, in which the catheter either failed before completion of 10 consecutive assembly/disassembly repetitions or the catheter completed 10 consecutive repetitions. The authors also conducted 100-repetition-maximum tests on 5 barium-impregnated and 5 rifampin-coated catheters using technique A.

Results: Catheter failure rates were significantly different among the different instruments used in assembly (p ≤ 0.001). Post hoc analysis showed that using mosquito forceps with shods resulted in a significantly lower catheter failure rate than the other instruments (p < 0.0005). The catheter failure rate of technique A was significantly lower than that of technique B (5% vs 81%, p < 0.001). There was no statistical difference between the failure rates of the barium and rifampin catheters (42% vs 44%, p = 0.9), but the barium catheters outperformed the rifampin catheters in the 100-repetition-maximum trials (p = 0.02). Instrument type (p = 0.0232) and catheter type (p = 0.0096) were both significant factors in determining the number of assembly/disassembly repetitions needed to cause catheter failure. It took an average of 2.79 repetitions of assembly/disassembly to cause catheter failure. DeBakey forceps had significantly lower mean repetitions to failure (mean 1.38) than the Gerald forceps without teeth (mean 2.73, p = 0.05) and bayonet (mean 3.25, p = 0.02).

Conclusions: This study was the first of its kind to analyze how physical manipulation affects the mechanical integrity of ventriculoperitoneal shunt components. The authors demonstrated a significantly lower shunt catheter failure rate when mosquito forceps with shods and technique A were used in assembly. Moreover, the authors established that use of DeBakey forceps resulted in failure with fewer assembly/disassembly repetitions than use of the Gerald and bayonet forceps, suggesting that DeBakey forceps are the most damaging instrument.

Download full-text PDF

Source
http://dx.doi.org/10.3171/2022.8.JNS22746DOI Listing

Publication Analysis

Top Keywords

catheter failure
24
surgical instruments
16
catheter
12
ventriculoperitoneal shunt
12
assembly/disassembly repetitions
12
failure rate
12
debakey forceps
12
failure
10
shunt assembly
8
surgical techniques
8

Similar Publications

We describe a 54-year-old man with type 2 diabetes mellitus, ischemic myopathy, pulmonary hypertension, and end-stage renal disease who was admitted for heart failure and listed for a dual cardiac-renal transplantation. Extensive calcification in the iliac arteries prevented clamping. Proximal endovascular balloon control of the left iliac artery was achieved using contralateral access; distal control was established by passing a Fogarty catheter distally through an iliac arteriotomy, later used for anastomosis of the cadaveric conduit.

View Article and Find Full Text PDF

Congenital heart disease (CHD) is the most common congenital anomaly in newborns. Advances in catheter and surgical techniques led to the majority of these patients surviving into adulthood, leading to evolving challenges due to the emergence of long-term complications such as arrhythmias. Interventional electrophysiology (EP) has had remarkable advances over the last few decades, and various techniques and devices have been explored to treat adult patients with CHD.

View Article and Find Full Text PDF

Objectives: A rising incidence of septic shock as well as recommendations for early vasopressor initiation has increased the number of patients eligible for norepinephrine (NE). Traditionally, NE has been administered through central lines, in intensive care units, due to the risk of extravasation in peripheral lines. The aim of the current study is to determine the rate of complications and patient outcomes when NE is administered through midline catheters (MCs) in intermediary care units (IMCUs).

View Article and Find Full Text PDF

Expansion of home hemodialysis (HHD) provides an opportunity to improve clinical outcomes, reduce cost of care, and address the staffing challenges currently faced in caring for patients with kidney failure on replacement therapy. To increase HHD expansion, current practices and barriers to home dialysis must be examined and addressed. One such barrier is vascular access for HHD; although tunneled hemodialysis central venous catheters (CVCs) have been used for decades, physicians still hesitate to send patients home without a mature, functional arteriovenous access.

View Article and Find Full Text PDF

Introduction And Aim:  Benign prostatic hyperplasia (BPH) is the enlargement and overgrowth of the prostate leading to the compression of the urethra and resulting in obstruction to the outflow of urine. Prostatic urethral lift (UroLift) is a budding minimally invasive technique that utilises mechanical manipulation of the prostate tissue so that the urethra is free from compression, thereby creating a channel for the outflow of urine. The aim of the audit was to assess the short- to medium-term outcomes in our centre in terms of improvement in symptoms, quality of life (QoL) and complication rates.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!