Introduction: Variability of the bipolar atrial electrogram amplitude may affect voltage maps created during ablation procedures, and thus also the extent of ablations. Therefore, the aim of the study was to assess the beat-to-beat electrogram amplitude variability in the left atrium in patients undergoing atrial fibrillation ablation.

Methods: In 11 patients undergoing ablation for atrial fibrillation, 362 mapping points were collected in two series. At each point, three consecutive beats were recorded and verified including the bipolar electrogram amplitude, contact force (CF), and orientation of the catheter tip. The repeatability and reproducibility of obtained measurements between consecutive beats and series were assessed by the Pearson correlation coefficient (r), the Bland-Altman test, repeatability coefficient (RC), relative standard deviation (RSD), and concordance correlation coefficient (CCC).

Results: A total of 1086 beats were analyzed. The correlation coefficient for bipolar atrial electrogram amplitude for the first two beats, and for the first and the third beats were 0.94 and 0.86, respectively. The average of differences between the first two beats and between the first and the third beats were 0.06 and 0.13 mV with 95% limits of agreement (LoA) within ±0.98 and ±1.74 mV, respectively. For CF values ≤5 and ≥20 g, the 95% LoA were narrower compared to other CF ranges and were ±0.49 and ±0.71 mV from the average value, respectively. When the analyzes were performed within the predefined ranges of bipolar electrogram amplitude: 0.05-1; 1-2; 2-3 mV, the 95% LoA were within ±0.33, ±0.98, and ±0.84 mV from the average value, respectively. RC and RSD were 1.41 mV and 20.8%, respectively. For repeated measurement between series, CCC ranged from 0.67 to 0.71 and the 95% LoA were within ±2.7 to 2.9 mV from the average value.

Conclusion: Bipolar atrial electrogram amplitude recorded at a given site during ablation procedures is variable to an extent that may be clinically relevant. The magnitude of the observed variability is greater during remapping.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jce.15702DOI Listing

Publication Analysis

Top Keywords

electrogram amplitude
28
atrial electrogram
16
atrial fibrillation
12
bipolar atrial
12
correlation coefficient
12
95% loa
12
amplitude variability
8
ablation procedures
8
patients undergoing
8
consecutive beats
8

Similar Publications

Correlation Between Voltage and Impedance Mapping in Patients with Atrial Fibrillation.

J Clin Med

December 2024

Department of Biomedical Sciences, Humanitas University, Via Rita Levi Montalcini 4, 20072 Pieve Emanuele, Italy.

Pulmonary vein isolation (PVI) represents the cornerstone of paroxysmal (PAF) and persistent atrial fibrillation (PsAF) ablation. Impedance values provide insights on tissue conductive properties. Consecutive patients undergoing PAF and PsAF ablation were prospectively enrolled.

View Article and Find Full Text PDF

High-density versus point-by-point electroanatomical activation mapping of a cranial vena cava tachycardia in a dog.

J Vet Cardiol

December 2024

Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, 930 Campus Road, Ithaca, NY, 14853, USA. Electronic address:

A seven-year-old, male intact Newfoundland was referred for catheter ablation of supraventricular tachycardia. Activation mapping was performed using an electroanatomical mapping system to visualize the activation wavefront in a color-coded fashion on an anatomical shell. Atrial flutter with an early-meets-late signal (i.

View Article and Find Full Text PDF

Introduction: Atrial pacing maps are often used as substitutes for sinus rhythm (SR) maps to expedite mapping procedures. However, the impact of this method on electrophysiological parameters has not been systematically examined. This study aimed to elucidate the advantages and limitations of atrial pacing maps.

View Article and Find Full Text PDF

Background: Electrical remodeling has been linked to the progression and recurrence of atrial fibrillation (AF) after catheter ablation (CA). Substrate mapping based solely on a voltage amplitude electrogram (EGM) does not provide a comprehensive understanding of the left atrial (LA) disease. The aim of this study is to assess left atrial spatial entropy (LASE) from voltage maps routinely obtained during AF ablation to further characterize the LA substrate.

View Article and Find Full Text PDF

Background: Cardiac magnetic resonance (CMR)-guided catheter ablation of the cavotricuspid isthmus (CTI) has been proven feasible, but determinants of local electrogram (EGM) voltage drops during radiofrequency (RF) applications are unknown.

Objective: The purpose of this study was to investigate local atrial bipolar EGM voltage drops and the association with delivered RF energy and anatomical information derived from peri-procedural CMR imaging.

Methods: In consecutive patients undergoing CMR-guided CTI ablation procedures, relative EGM voltage drops for RF applications ≥20 seconds were calculated.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!