Accurate and reproducible acetabular component positioning is among the most important technical factors affecting outcomes of total hip arthroplasty. Although several studies have investigated the influence of pelvic tilt and obliquity on functional acetabular anteversion, the effect of pelvic axial rotation has not yet been established. We analyzed a generic simulated pelvis created using preoperative full-body standing and sitting radiographs. A virtual acetabulum was placed in 144 different scenarios of acetabular anteversion and abduction angles. In each scenario, the effects of pelvic tilt and pelvic axial rotation on different combinations of acetabular orientations were assessed. The change in acetabular anteversion was 0.75° for each 1° of pelvic tilt and was most linear in abduction angles of 40°±45°. The change in acetabular anteversion was 0.8° for each 1° of pelvic axial rotation. Surgeons may consider adjusting acetabular anteversion in fixed axial pelvic deformities when the degree of deformity affects functional acetabular positioning, assessed from preoperative standing and sitting weight-bearing radiographs. [. 2023;46(1):e27-e30.].

Download full-text PDF

Source
http://dx.doi.org/10.3928/01477447-20221003-04DOI Listing

Publication Analysis

Top Keywords

acetabular anteversion
20
axial rotation
16
functional acetabular
12
pelvic tilt
12
pelvic axial
12
acetabular
9
pelvic
8
standing sitting
8
abduction angles
8
change acetabular
8