Purpose: In treatment planning for high-dose-rate (HDR) single-channel vaginal cylinder brachytherapy, dose distribution along the cylinder is influenced by the anisotropy of the source. Differences in anisotropy are due to differences in source dimensions and characteristics. In this study, we compared HDR vaginal cylinder brachytherapy treatment plans from two afterloader/treatment planning systems.

Material And Methods: Seventy-five plans with prescription to the surface were generated for cylinders in Varian BrachyVision and Elekta Oncentra. To understand the impact of source anisotropy on dose distribution to the surface of the cylinder, potential effect caused by differences in cylinder geometry between systems was eliminated by re-planning Varian cylinder using Elekta source model. Mean relative dose was calculated for each point as well as the dome and length of the cylinder. Related-samples Wilcoxon signed-rank tests were performed to compare the mean relative dose between systems.

Results: Treatment plans with VariSource iX source and cylinder demonstrated 16.2% lower ( < 0.001) dose at the tip compared to Elekta v.3. Average dose to the points along the dome of cylinder was 128.4% ±17.9% prescription dose with VariSource iX source and cylinder, and 99.9% ±4.3% with Elekta v.3 source and cylinder. For the same cylinder geometry, the effect of source characteristics produced up to 36.8% difference in dose homogeneity. When cylinder types were planned with the same source, there was no significant difference in dose distribution.

Conclusions: This study demonstrates that the effect of source characteristics produced up to 37% difference in dose homogeneity when comparing two afterloader/treatment planning systems, independent of cylinder geometry. This insight on variation in dose surrounding source system is imperative for dosimetry considerations. Depending on the choice of afterloader, the extent of EQD for tumor control versus normal tissue toxicity can vary.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9528830PMC
http://dx.doi.org/10.5114/jcb.2022.118633DOI Listing

Publication Analysis

Top Keywords

cylinder
14
cylinder geometry
12
source cylinder
12
difference dose
12
dose
11
source
11
treatment planning
8
planning systems
8
vaginal cylinder
8
cylinder brachytherapy
8

Similar Publications

A switching active disturbance rejection control (SADRC) strategy was proposed to solve the composite disturbance challenge arising from gap, LuGre friction, hydraulic spring force, and external load disturbance in the double closed-loop digital hydraulic cylinder position control system. Firstly, leveraging the established mathematical model of the double closed-loop digital hydraulic cylinder, the high-order state equation was derived. Subsequently, the high-order double closed-loop digital hydraulic cylinder control system was transformed into a second-order integral series control system using ADRC strategy.

View Article and Find Full Text PDF

Background: Inhalational anesthetic agents are a major source of potent greenhouse gases in the medical sector, and reducing their emissions is a readily addressable goal. Nitrous oxide (NO) has a long environmental half-life relative to carbon dioxide combined with a low clinical potency, leading to relatively large amounts of NO being stored in cryogenic tanks and H cylinders for use, increasing the chance of pollution through leaks. Building on previous findings, Stanford Health Care's (SHC's) NO emissions were analyzed at 2 campuses and targeted for waste reduction as a precursor to system-wide reductions.

View Article and Find Full Text PDF

Purpose: To evaluate the clinical outcome of laser-assisted surgical correction of high hyperopic or mixed astigmatism using small incision intrastromal lenticule rotation (SMILERO) alone or combined with photorefractive keratectomy (PRK).

Methods: This retrospective case series enrolled 25 eyes with high astigmatism that underwent SMILERO surgery. Uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), uncorrected near visual acuity (UNVA), manifest refraction, central corneal thickness, and corneal higher order aberrations were analyzed before surgery and after 3, 6, and 12 months of follow-up.

View Article and Find Full Text PDF

Purpose: To investigate long-term corneal biomechanical changes in thin corneas after small incision lenticule extraction (SMILE).

Methods: Patients with indications for SMILE were enrolled in this study between November 2017 and March 2018. Patients were matched for age, spherical diopter, cylinder, spherical equivalent (SE), and lenticule thickness (LT), and then categorized into the thin cornea group (preoperative thinnest central corneal thickness [CCT] of 500 µm or less, 32 eyes) or normal cornea group (CCT of greater than 500 µm, 32 eyes).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!