Purpose: To evaluate the reproducibility of magnitude of postoperative IOL rotation following implantation of a toric intraocular lens (IOL) with modified haptics, in comparison with a Proof-of-Concept (POC) study of prototype IOLs featuring the same haptic design.

Patients And Methods: A post-market, prospective, multicenter, single-arm, open-label clinical study was conducted. TECNIS Toric II IOL (Johnson & Johnson Vision, Irvine, CA, USA, Models ZCU150 to 600) were implanted in 125 subjects and evaluated at 1-day and 1-week postoperatively. An objective photographic method was used to determine postoperative IOL rotation. Uncorrected distance visual acuity (UCDVA), postoperative astigmatism, and surgeon satisfaction were also assessed. Rotation data were compared to the POC study in which two prototype non-toric monofocal IOLs, one with the same haptic design as Model ZCU, were studied.

Results: Mean absolute rotation was 0.82° ± 1.0° and 0.84° ± 0.92°at 1-day and 1-week visits, respectively. The percentage of eyes with ≤5° of absolute rotation was 98.9% and 99.5% at the 1-day and 1-week visits, respectively. The magnitude of rotation was similar to the POC study prototype IOLs. At 1-week, mean monocular UCDVA was 0.026 ± 0.135 (~20/21) logMAR and mean residual manifest refractive cylinder was 0.30 D ± 0.35 D. The mean signed axis difference (postoperative minus operative) of the TECNIS Toric II IOL was 0.23° ± 1.27° at 1-day and -0.07° ± 1.25° at 1-week, indicating a clockwise drift. At 1-week, surgeons were very satisfied or satisfied with overall clinical outcomes and rotational stability in 98% of implanted eyes.

Conclusion: The TECNIS Toric II IOL, with frosted, squared haptics, demonstrated low magnitude of postoperative IOL rotation, excellent uncorrected distance vision, and minimal residual astigmatism. The POC study design was supported, demonstrating that prototype non-toric monofocal IOLs can predict clinical performance of toric IOLs with the same haptic design.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9529011PMC
http://dx.doi.org/10.2147/OPTH.S373976DOI Listing

Publication Analysis

Top Keywords

poc study
16
postoperative iol
12
iol rotation
12
study prototype
12
tecnis toric
12
toric iol
12
1-day 1-week
12
reproducibility magnitude
8
rotation
8
rotation implantation
8

Similar Publications

Self-powered dual-photoelectrode photoelectrochemical aptasensor amplified by hemin/G-quadruplex-based DNAzyme.

Mikrochim Acta

January 2025

Department of General Surgery, Qilu Hospital (Qingdao), Cheeloo College of Medicine, Shandong University, 758 Hefei Road, Qingdao, Shandong, 266035, P.R. China.

A self-powered dual-electrode aptasensor was developed for the detection of tumor marker carcinoembryonic antigen (CEA). The composite BiVO/ZnInS, which is capable of forming a Z-scheme heterojunction, was chosen as the photoanode, and the AuNP/CuBiO complex was chosen as the photocathode in photoelectrochemical (PEC) detection. The experiments showed that the constructed self-powered dual-electrode system had a good photoelectric response to white light, and the photocurrent signal of the photocathode was significantly enhanced under the influence of the photoanode.

View Article and Find Full Text PDF

Continuous glucose monitoring (CGM) use in people with type 1 diabetes (T1D) is revolutionizing management. Use of CGM in hospital is poised to transform care, however routine use is not currently recommended due to lack of accuracy validation in acute care, including in people with T1D. We aimed to determine real-world CGM accuracy in hospitalized adults with T1D.

View Article and Find Full Text PDF

Glycemic Management of Patients with Hospital Hyperglycemia: A Retrospective Cohort Study on Adults Admitted in the Non-ICU Wards.

Diabetes Metab Syndr Obes

January 2025

Department of Endocrinology and Metabolism, Ningbo No.2 hospital, Ningbo, Zhejiang Province, People's Republic of China.

Purpose: To identify the key populations for Hospital Hyperglycemia (HH) management and to assess recent trends in the management of HH.

Patients And Methods: This retrospective study analyzed 1,136,092 point-of-care blood glucose (POC-BG) measurements from 40,758 patients with HH in non-intensive care unit (non-ICU) wards at Ningbo No.2 hospital from January 2020 to December 2022.

View Article and Find Full Text PDF

Background: The efficacy of immune checkpoint inhibitors (ICIs) depends on the tumor immune microenvironment (TIME), with a preference for a T cell-inflamed TIME. However, challenges in tissue-based assessments via biopsies have triggered the exploration of non-invasive alternatives, such as radiomics, to comprehensively evaluate TIME across diverse cancers. To address these challenges, we develop an ICI response signature by integrating radiomics with T cell-inflamed gene-expression profiles.

View Article and Find Full Text PDF

Background: There is a lack of data on the validation and diagnostic performance of the Freestyle Optium Neo-H (Freestyle) and Centrivet GK (Centrivet) devices for the diagnosis of hypoglycaemia, hyperglycaemia and hyperketonaemia in goats.

Objectives: The aim of the present study was to validate the Freestyle and Centrivet for the analysis of whole blood beta-hydroxybutyric acid (BHBA) and to validate the Freestyle for the analysis of whole blood glucose concentrations using the reference method (RM) in goat blood collected from the jugular and ear veins.

Methods: Venous blood samples were utilised to assess glucose and BHBA concentrations using the Freestyle, Centrivet and RM.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!