Background: A subcutaneous implantable cardioverter defibrillator (S-ICD) has several advantages over the transvenous ICD, including a reduced risk of lead-related mechanical complications and infection. However, inappropriate shock therapy is one of the most common adverse events associated with S-ICDs. We herein report a case of inappropriate shock therapy of S-ICD due to incomplete sealing of the seal plug.
Case Summary: A 60-year-old man, who had been on haemodialysis with a history of myocardial infarction, was transferred to the hospital after successfully being resuscitated from ventricular fibrillation (VF). An S-ICD was implanted for secondary prevention. On the third and the seventh post-operative days, S-ICD shock therapy was delivered without any tachyarrhythmias. As device interrogation revealed reproducible noises in both the secondary and alternate vectors by tapping at the generator, the sensing vector was fixed to the primary vector. Two months after discharge, the patient died of VF after receiving appropriate S-ICD shock delivery seven times. The S-ICD was retrieved from the body, and it was revealed that the seal plug had incompletely sealed and returned to its normal closed position after reinsertion of a torque wrench.
Discussion: Seal plug damage is a rare complication but should be considered if noise oversensing is provoked only at the secondary and/or alternate vectors. In the present case, the inappropriate shock therapy might have been prevented if we had checked the seal plug carefully. Therefore, we advocate confirming the seal plug routinely after the removal of the torque wrench.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9521653 | PMC |
http://dx.doi.org/10.1093/ehjcr/ytac366 | DOI Listing |
Adv Orthop
October 2024
3rd Department of Orthopedic Surgery, National & Kapodistrian University of Athens, KAT General Hospital, Athens, Greece.
BMJ Case Rep
November 2024
Vascular Surgery, Toronto General Hospital, Toronto, Ontario, Canada
Different techniques and devices can achieve haemostasis after endovascular procedures to reduce time to haemostasis and facilitate percutaneous large-bore access. One such device is the Angio-Seal, which accomplishes haemostasis by compressing a collagen plug against the arteriotomy. Multiple complications related to Angio-Seal exist, most commonly acute artery occlusions and rarely infections and endarteritis.
View Article and Find Full Text PDFEur Heart J
October 2024
Department of Cardiovascular Diseases, German Heart Center Munich, Technical University Munich University Hospital, Munich, Germany.
Background And Aims: Data from randomized trials investigating different access closure strategies after transfemoral transcatheter aortic valve implantation (TF-TAVI) remain scarce. In this study, two vascular closure device (VCD) strategies to achieve hemostasis after TF-TAVI were compared.
Methods: The ACCESS-TAVI (Comparison of Strategies for Vascular ACCESS Closure after Transcatheter Aortic Valve Implantation) is a prospective, multicenter trial in which patients undergoing TF-TAVI were randomly assigned to a strategy with a combined suture-/plug-based VCD strategy (suture/plug group) using one ProGlideTM/ProStyleTM (Abbott Vascular) and one Angio-Seal® (Terumo) versus a suture-based VCD strategy (suture-only group) using two ProGlidesTM/ProStylesTM.
BMC Oral Health
September 2024
Department of Conservative Dentistry & Endodontics, Saveetha Dental College & Hospitals, Saveetha Institute of Medical and Technical Sciences, Saveetha University, Chennai, 600077, India.
Background: Apexification is a procedure that promotes apical closure by forming mineralized tissue in the apex region of a nonvital young permanent tooth. Calcium silicate-based cement like Mineral trioxide aggregate (MTA) and Biodentine are commonly employed as apical barriers to facilitate this process. Microleakage, defined as the leakage along the junction between the canal wall and filling material, is a crucial aspect to assess in MTA and Biodentine applications as apical barriers, as it directly impacts the prevention of bacterial seepage and maintenance of structural integrity.
View Article and Find Full Text PDFStomatologiia (Mosk)
August 2024
Central Research Institute of Dentistry and Maxillofacial Surgery, Moscow, Russia.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!