Objective:  The aim of this study was to compare the cyclic fatigue of heat-treated and non-treated instruments.

Materials And Methods:  Twenty instruments with and without heat treatment from Bassi Logic were evaluated ( = 10). All instruments were subjected to dynamic cyclic fatigue through continuous rotations inside a stainless-steel tube (1.4-mm diameter, 9-mm curvature with 6-mm radius, and 90° angle) using a custom-made device, which performed 1 axial oscillation every 2 seconds with an amplitude of 3 mm, powered by a torque-controlled motor (Silver Reciproc, VDW, Germany), with speed adjusting to 950 rpm and torque to 4 N, according to manufacturer's guidance. The groups were compared using the Mann-Whitney test.

Result:  The fracture time of the heat-treated instruments (97.20 ± 39.94 second and non-treated instruments (14.30 ± 6.78 seconds had statistical differences [ < 0.05]). Heat-treated instruments were 6.8 times more resistant to fatigue than non-treated instruments ( < 0.05).

Conclusion:  Heat treatment provides increased fatigue resistance of NiTi alloy with the same design.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10329553PMC
http://dx.doi.org/10.1055/s-0042-1747952DOI Listing

Publication Analysis

Top Keywords

cyclic fatigue
12
heat treatment
8
instruments
5
influence heat
4
treatment nickel-titanium
4
nickel-titanium instruments
4
instruments cyclic
4
fatigue resistance
4
resistance simulated
4
simulated curved
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!