The pulmonary artery catheter (PAC)-despite its invasiveness-remains the gold standard for cardiac output (CO) monitoring. The FloTrac system, a less invasive hemodynamic monitor has been developed, which estimates CO using arterial pressure waveform analysis without external calibration. Recently, an upgraded version of FloTrac system with improved algorithm to follow changes in vascular resistance was introduced into the market. The aim of this study was to assess the reliability of the CO estimated from the fourth-generation FloTrac/EV1000 system (CO) compared to that measured with PAC using the thermodilution method (CO) during robotic-assisted off-pump coronary artery bypass (OPCAB) surgery. CO and CO were obtained simultaneously at 4 predefined time points during robotic-assisted OPCAB: 5 min after the induction of general anesthesia (T1), after starting one-lung ventilation (T2), after capnothorax (T3), and after mini-thoracotomy was performed (T4). The agreement of data was investigated by Bland-Altman analysis. Thirty-four patients were initially enrolled. After exclusion, 32 patients and a total of 128 paired CO measurements were obtained. The overall bias was 1.46 L/min, the 95% limits of agreements were - 3.40 to 6.33 L/min, and the percentage error was 72.98%. Regression analysis of the systemic vascular resistance index (SVRI) and the bias between CO and CO showed that the bias was moderately correlated with the SVRI (r = 0.43; p < 0.0001). Despite a software upgrade, the reliability of the fourth-generation FloTrac/EV1000™ system during robotic-assisted OPCAB to estimate CO was not acceptable, especially in patients with low SVRI.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00380-022-02177-xDOI Listing

Publication Analysis

Top Keywords

cardiac output
8
robotic-assisted off-pump
8
off-pump coronary
8
coronary artery
8
artery bypass
8
flotrac system
8
vascular resistance
8
validation cardiac
4
output estimation
4
estimation fourth-generation
4

Similar Publications

Purpose: Catheter ablation (CA) for atrial fibrillation (AF) in heart failure patients with preserved ejection fraction (HFPEF) has shown promising results in reducing mortality and improving heart function. However, previous studies have been limited by a lack of control groups and significant heterogeneity in their methodologies.

Hypothesis: CA for AF in HFPEF patients may not increase the complications and had similarly the rate of freedom from AF vs.

View Article and Find Full Text PDF

Background: ST-elevation myocardial infarction (STEMI) demands near-time reperfusion to reduce the risk of long-term heart failure. This study evaluates the proportion of impaired left ventricular ejection fraction (LVEF) following STEMI in the context of current healthcare settings at a tertiary care center equipped with the most advanced and up-to-date standards of care.

Methods: Patients experiencing STEMI as their first manifestation of coronary artery disease were analyzed, as these individuals had no prior experience with heart-related chest pain.

View Article and Find Full Text PDF

Background: Evidence-based beta-blockers are essential in managing heart failure with reduced ejection fraction (HFrEF) and are known to improve cardiovascular outcomes. Despite robust nascent guideline recommendations, studies indicate that beta-blockers are often underutilized or administered below target doses. This shivery issue is particularly relevant in Ethiopia, where comprehensive evaluations of beta-blocker utilization and dosing practices are limited.

View Article and Find Full Text PDF

Under low O, the heart of Carassius auratus (goldfish) shows an enhanced hemodynamics. This is observed in ex vivo cardiac preparations from animals acclimated to both normoxia and short-term (4 days) moderate hypoxia and perfused for 90 min with a hypoxic medium. Under short-term hypoxia, this is associated with a higher ventricular muscularity and an expanded mitochondrial compartment.

View Article and Find Full Text PDF

The microcirculation : master in normal pregnancy, puppet in preeclampsia.

Am J Obstet Gynecol

January 2025

Hasselt University, Faculty of Medicine and Life Sciences, Agoralaan, 3590 Diepenbeek, Belgium; Department of Obstetrics and Gynaecology, ZOL Genk, campus St. Jan, Schiepse Bos 6, 3600 Genk, Belgium.

Objective: To investigate the association between functional parameters of the microcirculation and the systemic cardiovascular system in a population of pregnant women at risk for gestational hypertension disorders.

Study Design: For this observational study, women at high cardiovascular risk according to maternal anthropometrics, obstetric and medical history, were recruited at random gestational age depending on time of referral to the outpatient clinic for high risk prenatal care at Ziekenhuis Oost Limburg, Genk Belgium. After birth, data of maternal and neonatal outcome were obtained from the hospital records: only women with normal pregnancy (n = 142) and with preeclampsia (n = 34) were included in this analysis.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!