Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: Thermal ablation of large tumors may benefit from simultaneous placement of multiple needles, but accurate placement becomes challenging as the number of needles increases. The aim of this work was to evaluate use of personalized needle guidance grid templates based on intraprocedural CT and fabricated at the point of care to implement ablation treatment plans with multiple needles in vivo.
Methods: A plastic frame was designed to hold two parallel plastic guide plates in a rigid relationship, fixed over the abdomen by a mounting arm. Steel ball targets (1.5 mm) were implanted under ultrasound in the livers of two domestic swine under general anesthesia. Following breath-hold CT of the subject and frame, the targets and frame were identified using customized 3D Slicer-based planning software. Multiple needle trajectories targeting the balls were planned, including complex off-plane trajectories. A machining program for drilling the hole pattern corresponding to the planned needle trajectories was generated. The pattern was drilled in the two plates with a numerical-controlled milling machine in the suite. The plates were attached to the frame and needles passed through the paired holes to the calculated depth. Placement accuracy was defined as needle tip-to-target distance on post-placement CT.
Results: The planning process and manufacturing required approximately 6 and 15 min, respectively. Needles were rapidly inserted (n = 11) to the target points without complications or traversing nontarget anatomy. The mean needle tip-to-target distance error was 3.4 ± 2.2, range 0-7 mm.
Conclusion: Rapid and accurate needle placement was feasible using a subject-specific, custom-drilled, needle guidance grid template fabricated intraprocedurally. Targeting accuracy and performance were similar to more complex and expensive tracking systems which may enable accurate intraprocedural implementation of treatment plans in the liver or other organs. This may be of value in complex ablation cases or in areas where more advanced guidance systems are not available.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9974549 | PMC |
http://dx.doi.org/10.1007/s11548-022-02747-4 | DOI Listing |