Improved quantitation of short-chain carboxylic acids in human biofluids using 3-nitrophenylhydrazine derivatization and liquid chromatography with tandem mass spectrometry (LC-MS/MS).

J Pharm Biomed Anal

Section of Nutrition Research, Division of Digestive Diseases, Department of Metabolism, Digestion and Reproduction, Faculty of Medicine, Imperial College London, Sir Alexander Fleming Building, South Kensington Campus, London SW7 2AZ, United Kingdom. Electronic address:

Published: November 2022

Short-chain carboxylic acids (SCCAs) produced by gut microbial fermentation may reflect gastrointestinal health. Their concentrations in serum and urine are indicative of specific metabolic pathway activity; therefore, accurate quantitation of SCCAs in different biofluids is desirable. However, it is often challenging to quantitate SCCAs since matrix effects, induced by the presence of a vast variety of other compounds other than SCCAs in complex biofluids, can suppress or enhance signals. Materials used for sample preparation may introduce further analytical challenges. This study reports for the first time a LC-MS/MS-based method to quantitate ten SCCAs (lactate, acetate, 2-hydroxybutyrate, propionate, isobutyrate, butyrate, 2-methylbutyrate, isovalerate, valerate and hexanoate) and evaluates the matrix effects in five human biofluids: serum, urine, stool, and contents from the duodenum and intestinal stoma bags. The optimized method, using 3-Nitrophenylhydrazone as a derivatization agent and a Charge Surface Hybrid reverse phase column, showed clear separation for all SCCAs at a concentration range of 0.1-100 µM, in a 10.5 min run without carry-over effects. The validation of the method showed a good linearity (R2 > 0.99), repeatability (CV ≤ 15%) assessed by intra- and inter-day monitoring. The lowest limit of detection (LLOD) was 25 nM and lowest limit of quantitation (LLOQ) was 50 nM for nine SCCA except acetate at 0.5 and 1 µM, respectively. Quantitative accuracy in all biofluids for most compounds was < ±15%. In summary, this methodology has the advantages over other techniques for its simple and fast sample preparation and a high level of selectivity, repeatability and robustness for SCCA quantification. It also reduced interferences from the matrix or sample containers, making it ideal for use in high-throughput analyses of biofluid samples from large-scale studies.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jpba.2022.115060DOI Listing

Publication Analysis

Top Keywords

short-chain carboxylic
8
carboxylic acids
8
human biofluids
8
serum urine
8
matrix effects
8
lowest limit
8
sccas
6
biofluids
5
improved quantitation
4
quantitation short-chain
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!