Purpose: To estimate the power of an implanted intraocular lens (IOL) by measuring IOL thickness using anterior segment optical coherence tomography (AS-OCT) and to assess the repeatability of measurements.
Methods: Ninety-seven eyes were studied one month after uneventful phacoemulsification within the bag Acrysof SA60AT IOL implantation (range +11 to +35). All eyes had postoperative refraction of 0.5 D of target refraction. AS-OCT was used to measure the central thickness of the IOL. Correlation between labelled IOL power and central IOL thickness as well as the measure of repeatability, for example, intraclass correlation coefficient (ICC), were evaluated. IOL thicknesses were also calculated using a formula and compared with AS-OCT derived measurements.
Results: IOL thickness correlated significantly with labelled IOL power (R = 0.985, 0.001). The regression equation (IOL Power = [0.04 IOL thickness in micron] - 7.56) indicates 25 microns of central IOL thickness change per 1D power change. Over the studied range, IOL power could be estimated with a precision of 0.85 0.02 D (95% confidence interval: 0.83-0.94D). ICC for repeated measurements was 0.999. There was a significant correlation between calculated and measured (AS-OCT) IOL thickness (R = 0.984, 0.001).
Conclusion: Central IOL thickness measurements with the AS-OCT are highly repeatable and closely correlated with the labelled IOL power, which can predict the IOL power with 0.85 D from the actual power. This method can be helpful in cases of postoperative IOL surprise.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9493428 | PMC |
http://dx.doi.org/10.18502/jovr.v17i3.11572 | DOI Listing |
J Cataract Refract Surg
January 2025
Kepler University Hospital, University Clinic for Ophthalmology and Optometry.
Purpose: To comprehensively evaluate the tilt of the crystalline lens in phakic eyes and its relationship with optical biometry parameters.
Setting: The retrospective analysis was conducted at the Department for Ophthalmology and Optometry, Kepler University Clinic in Linz, Austria.
Design: Lens tilt assessment was performed through a combined analysis of optical biometry (IOL Master 700) and anterior segment optical coherence tomography (AS-OCT) data to understand its relationship.
Front Artif Intell
January 2025
School of Medicine, Stanford University, Palo Alto, CA, United States.
Given close relationships between ocular structure and ophthalmic disease, ocular biometry measurements (including axial length, lens thickness, anterior chamber depth, and keratometry values) may be leveraged as features in the prediction of eye diseases. However, ocular biometry measurements are often stored as PDFs rather than as structured data in electronic health records. Thus, time-consuming and laborious manual data entry is required for using biometry data as a disease predictor.
View Article and Find Full Text PDFGMS Ophthalmol Cases
December 2024
Department of Ophthalmology, Disha Eye Hospital, Siliguri, India.
Background: Pseudophakic cystoid macular edema (CME) following primary anterior-chamber intraocular lens (ACIOL) implantations is commonly seen. Intravitreal triamcinolone acetonide (IVTA) injections have shown significant improvement in visual acuity and retinal thickness in refractory pseudophakic CME. Pseudohypopyon following IVTA injection is a known entity.
View Article and Find Full Text PDFPLoS One
December 2024
Department of Ophthalmology, Samsung Medical Center School of Medicine, Sungkyunkwan University, Seoul, Republic of Korea.
Background: To evaluate the ocular biometry agreement and prediction of postoperative refractive outcomes obtained using two swept-source optical coherence tomography (SS-OCT) biometers: Anterion (Heidelberg Engineering, Heidelberg, Germany) and Argos (Alcon, Fort Worth, TX, USA).
Methods: Ambispective analysis was conducted on 105 eyes at the Samsung Medical Center, Seoul, Republic of Korea, between June 2021 and March 2022. Biometric values were assessed using both devices before cataract surgery.
Acta Ophthalmol
December 2024
Department of Ophthalmology and Vision Science, Eye and ENT Hospital, Fudan University, Shanghai, China.
Purpose: To evaluate the long-term effects of implantable collamer lens (ICL) V4c on the biomechanical and morphological properties of crystalline lenses in Chinese adults with high myopia using Brillouin microscopy (BM).
Methods: In this case-control study, 102 highly myopic eyes from 102 patients (18-48 years, preoperative spherical equivalent (SE): -6.00--21.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!