Unlabelled: This multicenter prospective cohort study assessed the safety and performance of the Trabecular Metal Total Ankle System (TM Ankle; Zimmer) for primary total ankle arthroplasty (TAA).
Methods: One hundred and twenty-one consecutive patients qualifying for primary TAA were enrolled in the study. All patients received the TM Ankle implant. Clinical outcome examinations and radiographic evaluations were conducted at 6 weeks, 6 months, 1 year, 2 years, and 3 years. Patient-reported outcome measures (PROMs) were evaluated with use of the EuroQol-5 Dimensions questionnaire (EQ-5D), Ankle Osteoarthritis Scale (AOS), American Orthopaedic Foot & Ankle Society questionnaire (AOFAS), and patient satisfaction at each time point. Complications were classified according to the Canadian Orthopaedic Foot and Ankle Society (COFAS) system.
Results: The average AOFAS, EQ-5D, AOS pain, and AOS difficulty scores showed significant improvement at 6 weeks, 6 months, 1 year, 2 years, and 3 years as compared with the preoperative baseline (p < 0.001). The Kaplan-Meier survival estimate for revision when used in primary cases was 97.35% at 3 years. During the 3 years of follow-up, 9 patients showed abnormal radiographic findings. Two ankles had intraoperative complications, 38 had complications that were non-surgical or device-related, and 3 ankles underwent revision.
Conclusions: The results of the present study indicated that patient well-being significantly increased following TAA with use of the TM Ankle. Radiographic parameters also demonstrated a low incidence of abnormal findings.
Level Of Evidence: Prognostic Level III. See Instructions for Authors for a complete description of levels of evidence.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9489156 | PMC |
http://dx.doi.org/10.2106/JBJS.OA.21.00143 | DOI Listing |
Comput Methods Biomech Biomed Engin
January 2025
Key Laboratory of Advanced Design and Simulation Techniques for Special Equipment, Ministry of Education, Hunan University, Changsha, China.
Total talus replacement has been demonstrated to increase ankle instability. However, no studies have explored how to enhance postoperative stability. This study aims to explore the effect of collateral ligament reconstruction on ankle stability by finite element analysis.
View Article and Find Full Text PDFBackground: Total ankle replacement (TAR) has evolved in the last decade from a procedure rife with complication and failure to a promising alternative to arthrodesis. The ability to maintain ankle joint range of motion is showing great promise in patient-reported outcomes, postsurgical pain, as well as long-term sequalae of joint fusion. Although TAR can be performed via either an anterior or lateral approach both with their own sets of benefits and potential complications, the consensus seems to be that one is no better than the other when performed by high-volume surgeons.
View Article and Find Full Text PDFSensors (Basel)
January 2025
Department of Electrical Engineering and Information Technology, University of Naples Federico II, 80125 Naples, Italy.
Parkinson's disease (PD) is characterized by a slow, short-stepping, shuffling gait pattern caused by a combination of motor control limitations due to a reduction in dopaminergic neurons. Gait disorders are indicators of global health, cognitive status, and risk of falls and increase with disease progression. Therefore, the use of quantitative information on the gait mechanisms of PD patients is a promising approach, particularly for monitoring gait disorders and potentially informing therapeutic interventions, though it is not yet a well-established tool for early diagnosis or direct assessment of disease progression.
View Article and Find Full Text PDFJ Clin Med
January 2025
Department of Biomedical and Neuromotor Sciences (DIBINEM), Alma Mater Studiorum University of Bologna, 40123 Bologna, Italy.
Pseudotumors are defined as exuberant non-neoplastic inflammatory masses. This condition can be associated with hip and knee arthroplasty but has not been reported in Total Ankle Arthroplasty (TAA). This paper reports a pseudotumor that formed following TAA, highlighting its clinical presentation, management, and histopathology.
View Article and Find Full Text PDFJ Clin Med
January 2025
Department of Mechanical Engineering, Yonsei University, Seoul 03722, Republic of Korea.
: Patient-specific instrumentation (PSI) in total knee arthroplasty (TKA) uses preoperative three-dimensional imaging to create cutting blocks tailored to patient anatomy. However, there is debate regarding the additional benefits of PSI in terms of improved alignment or functional outcomes compared to using conventional instruments. Although PSI design has undergone continuous development, the improvements have not been incorporated.