Strength of titanium-zirconium alloy implants with a conical connection after implantoplasty.

J Prosthet Dent

Associate Professor and Director, Advanced Prosthodontics Program, Division of Restorative and Prosthetic Dentistry, The Ohio State University College of Dentistry, Columbus, Ohio. Electronic address:

Published: September 2024

Statement Of Problem: Peri-implantitis occurs around dental implants, and implantoplasty has been used to address this ongoing disease; however, the changes to the physical properties of an implant after implantoplasty have not been well documented.

Purpose: The purpose of this in vitro study was to determine the effect of implantoplasty on fracture strength and the load required for plastic deformation after cyclic fatigue on dental implants.

Material And Methods: Twenty-six titanium/zirconium (TiZr) alloy implants (Roxolid Bone Level Implant; 4.1×10 mm) were embedded with 50% thread exposure and divided into 4 groups based on whether they had implantoplasty treatment by using different diamond rotary instruments and/or cyclic loading at 250 N for 2 million cycles: C0 (control, no cyclic loading), T0 (test, no cyclic loading), CM (control, cyclic loading), and TM (test, cyclic loading). After implantoplasty and/or cyclic loading, all implants underwent a load-to-failure test. The maximum fracture strength (FS) and load required for the onset of plastic deformation (PD) were recorded in Newtons. One-way ANOVA and nonparametric comparisons with control by using the Dunn and Wilcoxon method for joint ranking were used for statistical analysis.

Results: The mean ±standard deviation FS for C0, CM, T0, and TM was 1465.2 ±86.4 N, 1480.7 ±64.1 N, 1299.3 ±123.8 N, and 1252.1 ±85.7 N, respectively. The mean ±standard deviation load for onset of PD for C0, CM, T0, and TM was 860.2 ±88.1 N, 797.0 ±130.5 N, 776.5 ±181.8 N, and 631.3 ±84.5 N, respectively. The TM group had a significantly lower FS and PD than the C0, CM, and T0 groups (P<.05) CONCLUSIONS: Both fracture strength (FS) and the onset of plastic deformation (PD) were significantly reduced after a TiZr alloy implant received implantoplasty and cyclic loading.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.prosdent.2022.08.015DOI Listing

Publication Analysis

Top Keywords

cyclic loading
24
alloy implants
8
fracture strength
8
strength load
8
load required
8
plastic deformation
8
and/or cyclic
8
control cyclic
8
loading test
8
test cyclic
8

Similar Publications

In situ growth of defective ZIF-8 on TEMPO-oxidized cellulose nanofibrils for rapid response release of curcumin in food preservation.

Carbohydr Polym

March 2025

Liaoning Key Lab of Lignocellulose Chemistry and BioMaterials, Liaoning Collaborative Innovation Center for Lignocellulosic Biorefinery, College of Light Industry and Chemical Engineering, Dalian Polytechnic University, Dalian 116034, China. Electronic address:

Uncontrolled release of active agents in active packaging reduces antimicrobial efficacy, hindering the effective protection of perishable products from microbial infection. Herein, a novel defective engineering was proposed to design defective and hollow ZIF-8 structures grown on TEMPO oxidized cellulose nanofibrils (TOCNFs) and use them as fast-reacting nanocarriers for loading and controlled release curcumin (Cur) in sodium alginate (SA) active packaging systems (CZT-Cur-SA). By employing stable chelation between tannic acid (TA) and ZIF-8 zinc ions, the connections between zinc ions and imidazole ligands were severed to form a loose and hollow structure, which facilitates the rapid reaction and release of active ingredients triggered by pH changes in the microenvironment.

View Article and Find Full Text PDF

Purpose: To investigate how varying ferrule heights and the number of glass fiber posts affect fracture resistance and behavior of endodontically treated maxillary first premolars with substantial loss of tooth structure.

Materials And Methods: Twenty-four extracted endodontically treated human maxillary first premolars were divided into three groups (n = 8) based on ferrule height and post number. The groups were as follows: premolars of 2 mm ferrule height that were restored with single posts (control group), premolars of 0.

View Article and Find Full Text PDF

Purpose: To evaluate the maximal load to failure, cyclic displacement, stiffness, and modes of failure of onlay subpectoral biceps tenodesis with an intramedullary unicortical metal button (MB) versus an inlay, all-suture Caspari-Weber (CW) technique.

Methods: Sixteen matched paired human cadaveric proximal humeri were randomly allocated for subpectoral BT with either CW or MB using a high-strength suture (N = 16; 8 male, 8 female, mean age = 82.5 years, range 62-99 years).

View Article and Find Full Text PDF

Rubber is widely used in situations involving cyclic loads, and the influence of temperature on rubber properties is particularly pronounced under cyclic loading. In this study, mechanical property tests and crack propagation tests of carbon black-filled hydrogenated nitrile butadiene rubber were conducted at four different operating temperatures. Based on the results of the crack propagation tests, the temperature-dependent characteristics of the Paris-Erdogan parameters and strain energy density were clarified.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!