A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Increased Sensitivity in Proton Transfer Reaction Mass Spectrometry by Using a Novel Focusing Quadrupole Ion Funnel. | LitMetric

Increased Sensitivity in Proton Transfer Reaction Mass Spectrometry by Using a Novel Focusing Quadrupole Ion Funnel.

Anal Chem

Anhui Province Key Laboratory of Medical Physics and Technology, Institute of Health and Medical Technology, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei 230031, P. R. China.

Published: October 2022

AI Article Synopsis

  • A new drift tube design called the focusing quadrupole ion funnel (FQ-IF) enhances sensitivity in proton transfer reaction mass spectrometry (PTR-MS).
  • The FQ-IF features 20 layers of stainless steel electrodes that taper from a larger to smaller diameter, allowing it to operate in both direct current (DC) and ion funnel (IF) modes.
  • Experimental results indicate the FQ-IF significantly improves transmission efficiency and sensitivity, achieving up to 87.9 times better sensitivity and a 35.7 times lower limit of detection compared to traditional DC mode.

Article Abstract

Sensitivity enhancement in proton transfer reaction mass spectrometry (PTR-MS) is an important development direction. We developed a novel drift tube called a focusing quadrupole ion funnel (FQ-IF) for use in PTR-MS to improve the sensitivity. The FQ-IF consists of 20 layers of stainless steel electrodes, and each layer has 4 quarter rings. The first 6 layers have a constant inner hole diameter of 22 mm; the latter 14 layers taper the inner diameter down to 8 mm. The FQ-IF drift tube can also operate in the direct current (DC) mode (similar to a conventional drift tube) and ion funnel (IF) mode (similar to a conventional ion funnel drift tube) by changing the voltage loading method. The simulation results show that the transmission efficiency of the FQ-IF is significantly improved compared to that of the other two modes. Further experiments show that the product ions of limonene tend to convert into smaller / fragment ions at higher voltages for the DC and IF modes. However, unlike the DC and IF modes, the distribution of product ions is stable at higher voltages for the FQ-IF. In other words, a higher RF voltage for the FQ-IF will not increase the collision energy of ions. In addition, the improvements in sensitivity for the FQ-IF range from 13.8 to 87.9 times compared to the DC mode and from 1.7 to 4.8 times compared to the IF mode for the 12 test compounds. The improvements in the limit of detection (LOD) for the FQ-IF range from 2.7 to 35.7 times compared to the DC mode. The FQ-IF provides a valuable reference for improving the sensitivity of PTR-MS and other mass spectrometers.

Download full-text PDF

Source
http://dx.doi.org/10.1021/acs.analchem.2c01893DOI Listing

Publication Analysis

Top Keywords

ion funnel
16
drift tube
16
times compared
12
compared mode