Background: Dry transthoracic pericardiocentesis is challenging and carries the risk of right ventricle (RV) or coronary artery injury. The RV can usually control bleeding automatically. For example, most perforations of the RV caused by pacemaker leads are treated without open surgery. Thus, we performed a transvenous puncture of the RV for dry pericardiocentesis with the back end of a 0.014-inch percutaneous transluminal coronary angioplasty (PTCA) guidewire and a 1.8 Fr microcatheter.

Methods: The back end of a 0.014-inch PTCA guidewire within a 1.8 Fr microcatheter was used to transvenously punctured through the middle of the acute margin of the RV into the pericardial space in 12 Yorkshire swine and 5 beagles. PTCA balloons of different diameters were used to dilate the puncture holes for 15 min under anticoagulation in all the animals to assess the ability of the RV to control the bleeding. Then, for 3 days, the puncture hole was dilated by a 6 Fr catheter in 9 swine and 5 dogs.

Results: The puncture was successful in all the animals. After withdrawal of the 2.5-mm balloon or the 6 Fr catheter, none of the animals exhibited pericardial effusion, as observed by echocardiography. There was no sustained ventricular arrhythmia or other complications. All the animals survived.

Conclusion: Transvenous puncture of the right ventricle with the back end of a 0.014-inch PTCA guidewire and 1.8 Fr microcatheter may be feasible and have a good safety margin.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9485541PMC
http://dx.doi.org/10.3389/fcvm.2022.974601DOI Listing

Publication Analysis

Top Keywords

ptca guidewire
16
transvenous puncture
12
0014-inch ptca
12
guidewire microcatheter
12
dry pericardiocentesis
8
puncture ventricle
8
ventricle 0014-inch
8
control bleeding
8
puncture
6
ptca
5

Similar Publications

A global treatment algorithm was developed for the endovascular revascularization of femoropopliteal lesions and chronic total occlusions, aiming toward a more standardized approach to endovascular treatment in patients with peripheral artery disease. The following steps are proposed. 1) Evaluation of lesion morphology based on preprocedural imaging by Duplex sonography and intravenous ultrasound for selection of lesion preparation tools.

View Article and Find Full Text PDF

Complications Following Retrieval of an Embolized Guide Catheter Tip Following Left Main Bifurcation PCI.

JACC Case Rep

February 2025

Department of Cardiology, Belfast Health and Social Care Trust, Belfast, Northern Ireland.

A 59-year-old woman presented with non-ST-segment elevation myocardial infarction. Angiography showed severe 3-vessel coronary disease. Percutaneous coronary intervention was successfully performed.

View Article and Find Full Text PDF

Removal of an entrapped guidewire by excimer laser coronary angioplasty in patients with chronic total occlusion intervention.

BMC Cardiovasc Disord

January 2025

Department of Cardiology, Fuwai Hospital, National Center for Cardiovascular Diseases, Peking Union Medical College, Chinese Academy of Medical Sciences, Beijing, China.

Guidewire entrapment (GE) is a rare complication that warrants complex interventions or surgical procedures. Here, we report the removal of an entrapped guidewire using excimer laser coronary angioplasty (ELCA) in a case of chronic total occlusion (CTO). Plaque tissue trapped with the guidewire was also removed.

View Article and Find Full Text PDF

Background: Advancing the retrograde microcatheter (MC) into the antegrade guide catheter during retrograde chronic total occlusion (CTO) percutaneous coronary intervention (PCI) can be challenging or impossible, preventing guidewire externalization.

Objectives: To detail and evaluate all the techniques focused on wiring to achieve intubation of the distal tip of a microcatheter, balloon, or stent with an antegrade or retrograde guidewire, aiming to reduce complications by minimizing tension on fragile collaterals during externalization and enabling rapid antegrade conversion in various clinical scenarios.

Methods: We describe the two main techniques, tip-in and rendezvous, and their derivatives such a facilitated tip-in, manual MC-tip modification, tip-in the balloon, tip-in the stent, deep dive rendezvous, catch-it and antegrade microcatheter probing.

View Article and Find Full Text PDF
Article Synopsis
  • Stent dislodgement is a rare but difficult complication during heart procedures, and existing treatment methods are not very effective.
  • This study tested a new self-expanding basket (SEB) catheter designed to retrieve dislodged stents without causing additional injury or needing surgery.
  • In a small trial with 6 patients, the SEB catheter successfully retrieved all dislodged stents, with no safety issues reported, indicating it may be a promising solution, but more extensive research is needed.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!