: Clinical decision support systems are advocated to improve the quality and efficiency in healthcare. However, before implementation, validation of these systems needs to be performed. In this evaluation we tested our hypothesis that a computerized clinical decision support system can calculate the CHADS-VASc score just as well compared to manual calculation, or even better and more efficiently than manual calculation in patients with atrial rhythm disturbances. : In = 224 patents, we calculated the total CHADS-VASc score manually and by an automated clinical decision support system. We compared the automated clinical decision support system with manually calculation by physicians. : The interclass correlation between the automated clinical decision support system and manual calculation showed was 0.859 (0.611 and 0.931 95%-CI). Bland-Altman plot and linear regression analysis shows us a bias of -0.79 with limit of agreement (95%-CI) between 1.37 and -2.95 of the mean between our 2 measurements. The Cohen's kappa was 0.42. Retrospective analysis showed more human errors than algorithmic errors. Time it took to calculate the CHADS-VASc score was 11 s per patient in the automated clinical decision support system compared to 48 s per patient with the physician. : Our automated clinical decision support system is at least as good as manual calculation, may be more accurate and is more time efficient.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9500878PMC
http://dx.doi.org/10.3390/medicina58091269DOI Listing

Publication Analysis

Top Keywords

clinical decision
32
decision support
32
support system
28
automated clinical
24
manual calculation
16
calculate chads-vasc
12
chads-vasc score
12
clinical
8
decision
8
support
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!