Pullout strength of connected pins in hand surgery: Effect of pin diameter and insertion angle.

Orthop Traumatol Surg Res

ICube CNRS UMR7357, Strasbourg University, 2-4, rue Boussingault, 67000 Strasbourg, France; Department of Hand Surgery, Strasbourg University Hospitals, FMTS, 1, avenue Molière, 67200 Strasbourg, France. Electronic address:

Published: December 2022

Objective: The purpose of this study was to examine the effect of insertion angles on the pullout strength of connected pins using a synthetic model simulating a hand bone.

Material And Methods: The material consisted of Sawbones® (20 mm×20 mm × 60 mm), fixation pins secured to a connecting rod much like an external fixator, an electric drill (speed 1,290 rpm) and a tensile testing machine. The Sawbones® were drilled with different pin diameters (1.2 mm, 1.5 mm, and 1.8 mm) and insertion angles (100°, 110° and 120°). A vertical displacement of 1 mm/min was applied until the pins were extracted (maximum force).

Results: The pullout strength increased with the insertion angle of the connected pins. It also increased with their diameter. Regardless of the pin diameter, the load-displacement curve during the pullout test had 4 sections (peak 1, ascending slope, peak 2, descending slope) that corresponded to the combined frictional force and contact force between the pins and Sawbones®.

Discussion: Our study findings showed that, theoretically, for wrist or hand fractures treated with connected pins, the larger the diameter and insertion angle, the better the mechanical holding power of the pins.

Level Of Evidence: I, experimental study.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.otsr.2022.103411DOI Listing

Publication Analysis

Top Keywords

connected pins
16
pullout strength
12
insertion angle
12
strength connected
8
pin diameter
8
diameter insertion
8
insertion angles
8
pins
7
insertion
5
pullout
4

Similar Publications

Introduction: Intramedullary femoral nails (IMFNs) need to be removed for subsequent joint replacement, refracture, nonunion, or infection. The tools used to extract newer IMFNs may not be suited for removal of older implants, especially if broken. The purpose of this study was to describe a novel technique in femoral nail extraction when primary measures fail and a report on 6 cases where it was used.

View Article and Find Full Text PDF

Emergency breakaway pins (EBPs) have been widely used in aircraft, especially in the suspension connection between the engine device and the airfoil. Currently, the existing EBPs, which are made of metal materials, barely satisfy the lightweight requirement of the airplane industry. Thus, the construction of a novel EBP with quartz fiber reinforced phenolics is proposed in this study, and the shear response is examined experimentally using a double-sided shear test.

View Article and Find Full Text PDF

Introduction: Subtalar joint arthrodesis is primarily indicated for advanced osteoarthritis, hindfoot deformity, and/or instability. During the first 6-10 weeks after surgery, there is an intermediary structurally weaker state before complete bony fusion of the calcaneus and talus occurs. Loading of the foot can lead to mechanical stresses and relative movements in the former joint gap, which can impede the fusion process.

View Article and Find Full Text PDF

Background: Eosinophilic esophagitis (EoE) predominantly affects males across all ages; however, little is known about sex differences for other aspects of EoE.

Objective: To investigate associations between sex and clinical presentation, endoscopic features, treatment choice and response in EoE patients in real-world practice.

Methods: Cross-sectional analysis of the multicenter EoE CONNECT registry.

View Article and Find Full Text PDF

Case: A 34-year-old man with a distal tibia bone defect was treated with an intramedullary bone transport nail (Precice Bone Transport System, NuVasive). During planned removal after successful treatment, 7 separate subcomponents of the nail became disconnected and had to be separately removed using specialized instrumentation. This occurred despite adherence to the manufacturer's recommended technique for nail removal and in the absence of clinical or radiographic evidence of implant failure.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!