Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Introduction: The purpose of this study was to report our initial experience of using a large-bore (0.096″ inner diameter) access catheter in neurovascular interventions.
Methods: Data were retrospectively collected from 5 sites in the US for neurovascular procedures performed using a large-bore access catheter. The effectiveness outcome was technical success, defined as the access catheter's successfully reaching its target vessel without conversion to direct carotid puncture or to a smaller-bore access catheter and successfully completing the intended neurointervention.
Results: One hundred and thirteen procedures performed in 112 patients were included in this study. The mean age of the patients was 67.5 years (SD 16.2), and about half (49.1%) were female. The most common primary access sites were the femoral (64.6%) or radial (32.7%) artery. Challenging anatomic variations included severe vessel tortuosity (26/81, 32.1%), type II aortic arch (17/88, 19.3%), type III aortic arch (14/88, 15.9%), bovine arch (16/104, 15.4%), severe angle (<30°) between the subclavian and target vessel (11/74, 14.9%), and subclavian loop (7/79, 8.9%). The median access time to branch view was 18 min (IQR 11-28, N = 75). The technical success rate was 94.7%. Two dissections (1.8%) were related to the large-bore access catheter. Access site complications occurred in 2 patients (1.8%). Four additional symptomatic periprocedural complications not related to the large-bore access catheter occurred (7.1%).
Conclusion: For neurovascular interventions, a 0.096″ inner diameter access catheter could be used with both femoral and radial arterial approaches, had a high technical success rate, and had a low rate of periprocedural complications.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11310716 | PMC |
http://dx.doi.org/10.1177/15910199221127074 | DOI Listing |