Purpose: To evaluate in vivo the feasibility, safety, and paclitaxel (PTX) deliverability of a newly developed non-commercially available Paclitaxel-Coated Balloon (PCB) catheter in the swine healthy peripheral vein model.

Materials And Methods: In total 12 PCBs were deployed in 12 venous segments. Primary feasibility endpoint was the successful application of the devices to the veins of the animals. Primary efficacy endpoint was the determination of the drug content in the venous tissue at 24 h and 7 days after balloon expansion, as assessed by analysis of the vein tissue with High Performance Liquid Chromatography (HPLC) coupled with tandem mass spectrometry. Primary safety endpoint was freedom from any major adverse event. Secondary endpoint was the investigation of any independent factor affecting the primary endpoints.

Results: Paclitaxel was detected in five out of six tissue samples 24 h post-intervention and five out of six tissues at 7 days following the procedure (10 tissue samples out of 12). The mean weight of tissue that was examined was 0.20604 ± 0.29822 g (range: 1.02823-0.03377 g) and the mean PTX concentration detected was 8.4 ± 13.1 μg/g (range: 0-36.1 μg/g). The average drug content detected at 24 h (17.1 ± 17.1 μg/g) was numerically superior, but non-statistically significant, compared to 7 days (3.1 ± 3.6 μg/g). An average of 33.8% of the drug remained on the balloon after retrieval. According to the multiple linear regression analysis, there was no significant correlation between transition time, PTX remaining on the balloon, time of analysis (24 h/7 days) and PTX tissue concentration. No abnormalities were noted during autopsy.

Conclusion: The newly developed PCB successfully delivered within the healthy venous wall a dose of Paclitaxel that inhibits neointimal hyperplasia. No safety issues were raised at short-term follow-up.

Download full-text PDF

Source
http://dx.doi.org/10.1177/11297298221122115DOI Listing

Publication Analysis

Top Keywords

paclitaxel-coated balloon
8
feasibility safety
8
newly developed
8
drug content
8
tissue samples
8
tissue
6
balloon
5
evaluation paclitaxel-coated
4
balloon catheter
4
catheter vivo
4

Similar Publications

Introduction: Initial surgical revascularization has a recognized primary role in patients with critical limb-threatening ischemia with a high-quality great saphenous vein for conduit. However, approximately one-third of lower extremity vein grafts develop lesions threatening graft patency. Traditional treatments have limitations, highlighting the need for innovative solutions.

View Article and Find Full Text PDF

Paclitaxel can inhibit smooth muscle cell proliferation and migration, and reduce the risk of vascular restenosis after balloon dilation. Our study investigated the safety and efficacy of paclitaxel-coated balloon (PCB) treatment for diabetic subpatellar artery disease. In this study, 140 patients with diabetic subknee arterial disease treated in our hospital from January 2022 to December 2023 were selected as the study objects, and were divided into the control group (conventional balloon interventionization angioplasty) and the observation group (PCB interventionization angioplasty), with 70 cases in each group according to the differences in previous balloon interventionization.

View Article and Find Full Text PDF

Background: The aim of this study is to assess the safety and feasibility of intravascular lithotripsy (IVL) for the treatment of the common femoral artery (CFA).

Methods: We analyzed patients who received IVL treatment for CFA from September 2021 to April 2023. All patients included presented with chronic limb-threatening ischemia.

View Article and Find Full Text PDF

Drug-coated balloons (DCBs) represent a promising alternative to drug-eluting stents (DESs) by adopting a "leave-nothing-behind" approach, avoiding the long-term implantation of metallic materials associated with vessel damage, stent thrombosis, and restenosis. Although DCBs have historically been indicated for patients with in-stent restenosis or de novo small-vessel disease, recent studies indicate the potential for broader applications in acute coronary syndromes (ACSs). Trials comparing DCB and DES treatments show that DCBs yield comparable long-term outcomes, with some studies suggesting reduced rates of secondary endpoints such as cardiac death and non-fatal myocardial infarction among ACS patients treated with DCBs.

View Article and Find Full Text PDF