A survey of the incidence of defibrillator damage during double sequential external defibrillation for refractory ventricular fibrillation.

Resusc Plus

University of Toronto, Department of Family and Community Medicine, Division of Emergency Medicine, Toronto, ON, Canada.

Published: September 2022

Background: Double Sequential External Defibrillation (DSED) is a proposed treatment strategy for patients in refractory VF (RVF) during out-of-hospital cardiac arrest (OHCA). Defibrillator damage employing DSED is a theoretical concern expressed by defibrillator manufacturers yet the incidence of damage during resuscitation remains unknown.

Objective: We sought to explore the incidence of defibrillator damage employing DSED for RVF during OHCA.

Methods: We conducted a survey of EMS agencies, authors of previous publications, EMS medical directors, base hospital medical oversight groups, and defibrillator manufacturers to assess the incidence of defibrillator damage during DSED. Our survey focused on the frequency of DSED use, number of shocks used during DSED, technique used to employ DSED (simultaneous or sequential), and the incidence of defibrillator damage during DSED. We specifically targeted groups that were known to be using DSED in clinical practice.

Results: Our survey response rate was 50% (65/129): 61% (34/56) EMS medical directors, 60% (6/10) authors, 100% (8/8) base hospitals, 33% (1/3) defibrillator manufacturers, 31% (16/52) paramedic services. In our case-based analysis the overall incidence of defibrillator damage was 0.4%. The incidence of defibrillator damage based on total number of DSED shocks was estimated between 0.11% and 0.22%. All reported cases of defibrillator damage occurred using a simultaneous defibrillation technique.

Conclusion: When DSED is employed using either a sequential or simultaneous technique the rate of defibrillator damage appears to be exceedingly low. Further high-quality evidence is required to determine the impact of DSED on patient centered outcomes, but the incidence of defibrillator damage should not limit it use. Defibrillator damage should continue to be monitored in future trials and clinical practice.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9464949PMC
http://dx.doi.org/10.1016/j.resplu.2022.100287DOI Listing

Publication Analysis

Top Keywords

defibrillator damage
44
incidence defibrillator
28
defibrillator
14
damage
12
dsed
12
defibrillator manufacturers
12
double sequential
8
sequential external
8
external defibrillation
8
damage employing
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!