Advances in sensor technology have provided an opportunity to measure gait characteristics using body-worn inertial measurement units (IMUs). Whilst research investigating the validity of IMUs in reporting gait characteristics is extensive, research investigating the reliability of IMUs is limited. This study aimed to investigate the inter-session reliability of wireless IMU derived measures of gait (i.e., knee angle, range of motion) taking multiple test administrators into account. Fifteen healthy volunteers (43 ± 15 years) completed two visits. Within each visit, participants were required to perform two sets of 6 gait trials (6-metre walk tests). IMUs were placed on the participant in 7 locations on the lower limbs and waist. A different test administrator (n = 3) applied the IMUs at each set. At visit 2, this procedure was repeated with the same test administrators as visit 1. Kinematic measures of maximum angle (Knee_Max), minimum angle (Knee_Min), and range of motion (RoM) are reported for the left and right knee. The intraclass correlation coefficients (ICC), standard error of measurement (SEM) and minimum detectable change (MDC) are reported to determine IMU reliability. The results confirmed moderate to good inter-session reliability across all features (0.73-0.87). SEM values ranged from 1.21-3.32° and MDC values ranged from 3.37 - 9.21°. Therefore, IMUs appear to be a reliable method to determine inter-session gait characteristics across multiple test administrators.

Download full-text PDF

Source
http://dx.doi.org/10.1109/EMBC48229.2022.9871584DOI Listing

Publication Analysis

Top Keywords

gait characteristics
12
test administrators
12
reliability wireless
8
inter-session reliability
8
range motion
8
multiple test
8
values ranged
8
imus
6
gait
5
test-retest reliability
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!