Objectives: The long-term goal of this study is to investigate the efficacy of a novel, ultrasound-based technique called subharmonic-aided pressure estimation (SHAPE) to measure bladder pressure as a part of a cystometrogram (CMG) in a urodynamic test (ie, pressure-flow study). SHAPE is based on the principle that subharmonic emissions from ultrasound contrast microbubbles (MBs) decrease linearly with an increase in ambient pressure. We hypothesize that, using the SHAPE technique, we can measure voiding bladder pressure catheter-free. This is of importance because the CMG catheter, due to its space-occupying property and non-physiological effects, can undermine the reliability of the test during voiding and cause misdiagnosis. In this study, we tested this hypothesis and optimized the protocol in a controlled benchtop environment.

Materials And Methods: A bladder phantom was designed and built, capable of simulating clinically relevant bladder pressures. Laboratory-made lipid-shelled MBs (similar in composition to the commercial agent, DEFINITY) was diluted in 0.9% normal saline and infused into the bladder phantom using the CMG infusion system. A typical simulated CMG consists of 1 filling and 4 post-filling events. During CMG events, the bladder phantom is pressurized multiple times at different clinically relevant levels (small, medium, and large) to simulate bladder pressures. Simultaneous with pressurization, MB subharmonic signal was acquired. For each event, the change in MB subharmonic amplitude was correlated linearly with the change in bladder phantom pressure, and the SHAPE conversion factor (slope of the linear fit) was determined. In doing so, a specific signal processing technique (based on a small temporal window) was used to account for time-decay of MB subharmonic signal during a simulated CMG.

Results: A strong inverse linear relationship was found to exist between SHAPE and bladder phantom pressures for each of the CMG filling and post-filling events ( r2> 0.9, root mean square error < 0.3 dB, standard error <0.01 dB, and P < 0.001). SHAPE showed a transient behavior in measuring bladder phantom pressure. The SHAPE conversion factor (in dB/cm H 2 O) varied between filling and post-filling events, as well as by post-filling time. The magnitude of the SHAPE conversion factor tended to increase immediately after filling and then decreases with time.

Conclusions: Microbubble subharmonic emission is an excellent indicator of bladder phantom pressure variation. The strong correlation between SHAPE signal and bladder phantom pressure is indicative of the applicability of this method in measuring bladder pressure during a CMG. Our results suggest that different SHAPE conversion factors may be needed for different events during a CMG (ie, at different time points of a CMG). These findings will help us better protocolize this method for introduction into human subjects and allow us to take the next step toward developing a catheter-free voiding CMG using SHAPE.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RLI.0000000000000919DOI Listing

Publication Analysis

Top Keywords

bladder phantom
24
bladder
10
bladder pressure
8
clinically relevant
8
bladder pressures
8
filling post-filling
8
post-filling events
8
subharmonic signal
8
pressure
6
phantom
6

Similar Publications

Background: Huntington's disease (HD) is a rare neurodegenerative disorder caused by an expansion of the CAG trinucleotide repeat in the huntingtin gene which encodes the mutant huntingtin protein (mHTT) that is associated with HD-related neuropathophysiology. Noninvasive visualization of mHTT aggregates in the brain, with positron emission tomography (PET), will allow to reliably evaluate the efficacy of therapeutic interventions in HD. This study aimed to assess the radiation burden of [F]CHDI-650, a novel fluorinated mHTT radioligand, in humans based on both in vivo and ex vivo biodistribution in mice and subsequent determination of dosimetry for dosing in humans.

View Article and Find Full Text PDF

. The International Commission on Radiological Protection (ICRP) decided to develop pregnant-female reference computational phantoms, including the maternal and fetal phantoms, through its 2007 general recommendations. Acknowledging the advantages of the mesh geometry, the ICRP decided to develop the pregnant-female mesh-type reference computational phantoms (MRCPs) for 8, 10, 15, 20, 25, 30, 35, and 38 week fetal ages directly in the mesh format.

View Article and Find Full Text PDF

Comparison of fluorescence intensity of protoporphyrin IX as observed on the screen of different cystoscopic systems.

Photodiagnosis Photodyn Ther

November 2024

Department of Urology, Kochi Medical School, Kohasu, Oko, Nankoku, Kochi, Japan; Center for Photodynamic Medicine, Kochi Medical School, Kohasu, Oko, Nankoku, Kochi, Japan.

Background: Photodynamic diagnosis-assisted transurethral resection of bladder tumor (PDD-TURBT) is an important and useful intervention used for removing non-invasive bladder cancer (NMIBC). Several PDD devices can be used in clinical practice, but few reports have compared them.In this study, we examined the differences in detected fluorescence intensity for each PDD device.

View Article and Find Full Text PDF
Article Synopsis
  • This study developed an algorithm that creates a referenceable bladder map from standard cystoscopy videos, allowing diagnosis and monitoring of bladder carcinoma without needing advanced equipment.
  • The algorithm generates 2D bladder maps by stitching together frames from cystoscopy videos by matching surface features, and it was tested on both swine and archived clinical videos.
  • Results showed high performance with 93-99% of frames suitable for creating bladder maps, useful for recording pathology and treatment areas, especially in cases of recurrent bladder cancer or in low-resource environments.
View Article and Find Full Text PDF

Aims: Magnetic resonance imaging (MRI)-only radiotherapy has emerged as a solution to address registration errors that can lead to missed dose delivery. However, the presence of systemic geometric distortion (SGD) stemming from gradient nonlinearity (GNL) and inhomogeneity of the main magnetic field (B) necessitates consideration. This study aimed to quantitatively assess residual SGD in 1.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!