Background: The prosthetic screw fixes the prostheses to the implants. Upon osteointegration, the untightening of the prosthetic screw is the most common problem in oral rehabilitation with implants.

Objective: To study the deformation of the implant retaining hexagonal screw head.

Methods: This investigation used two methods for evaluating the screw head's area of deformation (mm ): a stereoscopic microscopy and micro computed tomography (microCT). For stereoscopic microscopy, 16 titanium alloy (T) and 16 titanium gold-plated alloy (G) screws of the Zimmer Biomet™ brand were used, divided into eight groups: group 0 (control group), groups T1 and G1 (screws tightened 10 times to 20 Ncm), the groups T2 and G2 (screws tightened 20 times to 20 Ncm) and the groups T3 and G3 (screws tightened 10 times to 30 Ncm). In the study with microCT, one screw was randomly chosen from each of the groups described above to perform the scanning by microCT.

Results: When comparing the type of screw material using stereoscopic microscopy, no statistically significant differences were found (p > 0.05). Contrarily, different number of successive grips and different torque value showed statistically significant differences in the head section of the retaining screws (p < 0.05). The observation by microCT showed the torque applied is crucial to the head deformation in titanium screws. In gold-plated screws the successive tightening appears to be pivotal.

Conclusion: Titanium and gold screws tend to behave similarly. By increasing the tightening cycles and the torque values of protocols greater levels of deformations can be expected. In general, microCT data showed better results for gold-plated titanium alloy.

Clinical Significance: To control severe screw head deformation and the impossibility of untightening the implant's restoration, clinicians should avoid extreme torque values and prevent surpassing 10 tightening cycles.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jerd.12959DOI Listing

Publication Analysis

Top Keywords

stereoscopic microscopy
16
groups screws
12
screws tightened
12
tightened times
12
deformation implant
8
implant retaining
8
prosthetic screw
8
times 20 ncm
8
20 ncm groups
8
statistically differences
8

Similar Publications

Distribution of microplastics in the soils of a petrochemical industrial region in China: Ecological and Human Health Risks.

Environ Geochem Health

December 2024

State Key Laboratory of Environmental Criteria and Risk Assessment, Chinese Research Academy of Environmental Sciences, Beijing, 100012, China.

Although microplastic pollution is a global concern, information on the distribution of microplastics in petroleum and petrochemical urban soils is limited. In this study, we investigated the occurrence, ecological risk, and human exposure risk of microplastics in different land-use types of soil in Daqing Administrative region, a prominent petroleum and petrochemical industrial base in China. Stereoscopic microscopy and Fourier transform infrared spectroscopy (μ-FTIR) were used to study the chemical composition and distribution characteristics of microplastics.

View Article and Find Full Text PDF

Especially, the processing and utilization of biomass-based material is closely related to the vessel, e.g. the flow of vapour and additive.

View Article and Find Full Text PDF

A solar photovoltaic (PV) cell, is an electrical device that uses the PV effect to convert light energy into electricity. The application of oyster mushroom dyes in dye sensitized solar cell (DSSC) is a novel strategy to substitute the costly chemical production process with easily extractable, environmentally acceptable dyes. Both dyes of yellow and pink oyster mushrooms were extracted using the same process but dried into powder form using two techniques, warm drying and freeze drying.

View Article and Find Full Text PDF

Purpose: Multi-zoom microscopic surface reconstructions of operating sites, especially in ENT surgeries, would allow multimodal image fusion for determining the amount of resected tissue, for recognizing critical structures, and novel tools for intraoperative quality assurance. State-of-the-art three-dimensional model creation of the surgical scene is challenged by the surgical environment, illumination, and the homogeneous structures of skin, muscle, bones, etc., that lack invariant features for stereo reconstruction.

View Article and Find Full Text PDF

This study explores the potential of zinc and silver nanocomposites, synthesized with β-lactoglobulin, a whey protein, in promoting wound healing, using the C57BL/6J mouse model. Our research is distinct in its dual focus: assessing the antimicrobial efficacy of these nanocomposites and their impact on wound healing processes. The antimicrobial properties were investigated through minimum inhibitory concentration (MIC) assessments and colony-forming unit (CFU) tests, providing insights into their effectiveness against wound-associated microorganisms.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!