Purpose: This study aimed to evaluate and compare the discriminating ability of corneal elevation maps generated using a swept-source optical coherence tomography (SS-OCT) (SS-OCT ANTERION, Heidelberg Engineering, Heidelberg, Germany), which was estimated with different reference surfaces, to distinguish normal corneas from those with keratoconus and keratoconus suspect.

Methods: A total of 126 eyes of patients, which comprised 43, 37, and 46 keratoconus, keratoconus suspects, and normal controls, respectively, were included in this study. The anterior and posterior elevations at the thinnest point under the best-fit sphere (BFS) and toric-ellipsoid (BFT), respectively, and other corneal parameters were measured using the SS-OCT. In addition, the receiver operating characteristic (ROC) curve analysis and cut-off value were calculated to evaluate the diagnostic ability of the corneal elevation values in differentiating keratoconus and keratoconus suspects from normal eyes.

Results: The mean total keratometric and corneal elevation values were significantly higher in the keratoconus group than in the other groups. Pachymetric parameters exhibited the lowest values for keratoconus. In addition, ROC curve analyses showed a high accuracy of the thinnest point anterior and posterior BFT for both keratoconus and keratoconus suspects and normal controls (area under the ROC were 0.969 and 0.961, respectively). Furthermore, the optimal cut-off point of the posterior elevation at the thinnest point under BFT was 16.44 μm (sensitivity and specificity of 86% and 98%, respectively) for differentiating keratoconus from normal and keratoconus suspect eyes.

Conclusions: The elevation map using the BFS and BFT references measured with the anterior segment SS-OCT is considered an effective indicator for keratoconus diagnosis. Therefore, the anterior segment SS-OCT can effectively differentiate keratoconus from suspected keratoconus and normal corneas by measuring parameters such as posterior and anterior elevations, pachymetry, and keratometry.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9436129PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0274071PLOS

Publication Analysis

Top Keywords

keratoconus
16
keratoconus keratoconus
16
anterior segment
12
corneal elevation
12
keratoconus suspects
12
suspects normal
12
thinnest point
12
normal keratoconus
8
swept-source optical
8
optical coherence
8

Similar Publications

Preparation for cataract surgery in patients with keratoconus (KC) is associated with particular challenges: calculating the optical power of the intraocular lens (IOL), selecting the appropriate IOL model, and considering additional interventions aimed at stabilizing KC and reducing irregular astigmatism. This article presents a fundamentally new approach to combined treatment of KC with cataract, consisting of two stages: first, performing bandage therapeutic-optical keratoplasty (BTOK), followed by cataract phacoemulsification as the second stage.

View Article and Find Full Text PDF

Objectives: To demonstrate corneal remodeling after corneal allograft intrastromal ring segment (CAIRS) with an anterior-segment optical coherence tomography (AS-OCT).

Design: A prospective observational single-center study.

Methods: This observational study included keratoconus patients who underwent CAIRS implantation into a stromal tunnel.

View Article and Find Full Text PDF

We have compared the biomechanical properties of human and porcine corneas using vibrational optical coherence tomography (VOCT). The elastic modulus of the cornea has been previously reported in the literature to vary from about several kPa to more than several GPa based on the results of different techniques. In addition, the formation of corneal cones near the central cornea in keratoconus has been observed in the clinic.

View Article and Find Full Text PDF

Keratoconus is a condition that causes progressive thinning and anterior protrusion of the cornea. Because of its irregular astigmatism, mild to moderate keratoconus is corrected with hard contact lenses (HCLs), but blepharoptosis due to the long-term wearing of HCLs is often a problem. In this study, we investigated blepharoptosis in HCL wearers with keratoconus.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!