Background: Continuous glucose monitors (CGMs) are widely used for individuals with diabetes mellitus, particularly those with type 1 diabetes (T1D). Advancements in CGM technology allow for glycemic assessment without capillary glucose measurements as many come factory calibrated. However, exercise, an essential component of diabetes care, has been reported to alter accuracy of earlier generation CGM. Considering the importance of physical activity for individuals with T1D and the progression of CGM technology, we aimed to investigate the accuracy of the Dexcom G6 during physical activity.

Methods: Adolescents (ages 13-20 years) exercised on a treadmill for 40 minutes, with a 10-minute break at minute 20. We obtained paired CGM and glucometer measurements before and every 10 minutes during and after exercise. Accuracy analysis was determined by mean absolute relative difference (MARD), mean absolute difference (MAD), and Clarke Error Grid Analyses.

Results: Mean absolute relative difference and MAD increased during exercise (14%-33% and 24.3-34 mg/dL) but improved after exercise. We noted certain CGM locations produced greater changes in accuracy as MARD and MAD increased markedly when the CGM was on the buttocks (18%-46% and 30-41 mg/dL). We also noted decreased odds of Zone A in the Clarke error grid when the CGM was on the buttocks compared to the abdomen (odds ratio [OR]: 0.146; = 0.0003; 95% CI = 0.052-0.415).

Conclusions: This CGM system showed alterations in accuracy during exercise. Our findings additionally suggest interstitial fluid changes in muscles during exercise alter accuracy of CGM; however, additional research is required.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11089875PMC
http://dx.doi.org/10.1177/19322968221120433DOI Listing

Publication Analysis

Top Keywords

cgm
9
continuous glucose
8
accuracy exercise
8
type diabetes
8
diabetes mellitus
8
cgm technology
8
alter accuracy
8
absolute relative
8
relative difference
8
difference mad
8

Similar Publications

Background: Both glucagon-like peptide 1 receptor agonists (GLP-1 RAs) and continuous glucose monitoring (CGM) have been shown to improve glycated hemoglobin A1c (A1c) levels among patients with type 2 diabetes mellitus (T2DM). Recently, a US real-world study found statistically significant improvements in A1c levels among patients using GLP-1 RA and a CGM device, compared with a matched cohort receiving only GLP-1 RA.

Objectives: To assess the cost-effectiveness from a US payer perspective of initiating CGM (FreeStyle Libre Systems) in people living with T2DM using a GLP-1 RA therapy, compared with GLP-1 RA alone.

View Article and Find Full Text PDF

Aims: To evaluate the use of hybrid (telehealth and in-person) care on visitation and glycaemia in older adults with type 1 diabetes (T1D).

Methods: In this retrospective study, we examined clinical characteristics, number of visits (telehealth and in-person) and continuous glucose monitoring (CGM) metrics for older adults (≥65 years) with T1D from electronic health records during the pre-COVID-19 pandemic (March 1, 2019-March 1, 2020; in-person) and pandemic (September 1, 2020-August 31, 2021; hybrid) periods. Main outcomes were the number of visits and changes in glycaemic control (HbA1c), and in a sub-group of older adults using CGM, changes in CGM metrics between in-person and hybrid care.

View Article and Find Full Text PDF

Background: Identification of genetic alleles associated with both Alzheimer's disease (AD) and concussion severity/recovery could help explain the association between concussion and elevated dementia risk. However, there has been little investigation into whether AD risk genes associate with concussion severity/recovery, and the limited findings are mixed.

Objective: We used AD polygenic risk scores (PRS) and APOE genotypes to investigate any such associations in the NCAA-DoD Grand Alliance CARE Consortium (CARE) dataset.

View Article and Find Full Text PDF

Background: Gestational Diabetes Mellitus (GDM) prevalence is rising worldwide, but optimal dietary strategies remain unclear. The eMOM pilot RCT compared a plant-protein rich Healthy Nordic Diet (HND) and a moderately carbohydrate restricted diet (MCRD) and their potential effects on time in glucose target range (≤ 7.8 mmol/L, %TIR), and on newborn body composition.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!