Carbon Dioxide, Oxygen, and Ammonia Levels in Mouse and Rat Disposable IVC Removed from Mechanical Ventilation.

J Am Assoc Lab Anim Sci

Pfizer Worldwide Research, Development and Medical, Comparative Medicine, Cambridge, Massachusetts.

Published: September 2022

Maintenance of an appropriate microenvironment for rodents used in research is of paramount importance because changes in environmental parameters such as O₂ and humidity can influence animal health and welfare and potentially alter research results. Here we evaluated the microenvironment of mouse and rat disposable cages after removal from mechanical ventilation in order to guide recommendations for their use. Cages with sealed IVC lids, unsealed lids (partially ajar), and lids without the exhaust filter (for rats) or static lids (for mice) were removed from the ventilated rack and were thereafter monitored CO₂, O₂, and NH₃ levels. For mice, effects were investigated under both standard (set point of 72°F/22°C) and thermoneutral (set point of 82°F/28°C) temperatures. When IVC with sealed lids and group-housed C57BL/6J male mice were removed from ventilation under standard temperatures, CO₂ started at 6,600 ± 265 ppm at 0 h and rose to 42,500 ± 7,263 ppm at 1 h, with mice showing a visibly elevated respiratory rate in 1 of the 3 cages; CO₂ stabilized at 26,150 ± 3,323 ppm at 8 h. In contrast, CO₂ levels in cages with single mice were stable after 1 h (1,350 ± 409 ppm at 0 h, 9,367 ± 802 ppm at 1 h, and 8,333 ± 1,115 ppm at 8 h). Findings were similar at thermoneutral temperatures: sealed group-housed mice cages started at 3,617 ± 475 ppm at 0 h and rose to 39,333 ± at 5,058 ppm at 1 h, whereas sealed cages with 1 mouse started at 1,117 ± 247 ppm at 0 h and were 7,500 ± 1,997 ppm at 8 h. IVC with sealed lids and pair-housed Crl:CD(SD) female rats rose to 48,000 ± 2,828 ppm CO₂ and over 70% humidity within 1 h. By 3 h, IVC with sealed lids and singly housed rats had 40,167 ± 5,132 ppm CO₂, and rats were displaying a visually elevated respiratory rate. O₂ levels had an inverse relationship with CO₂ levels. Removing the rat lid exhaust filter was not helpful. However, leaving the IVC lid ajar ameliorated the rise in CO₂ and fall in O₂ for both species. Therefore, IVC with sealed lids and group-housed mice should not be removed from ventilation more than 1 to 2 h; IVC containing pair- or singly-housed rats IVC should not be removed for more than 1 or 3 h, respectively. Whenever possible, such cages should be fitted with static lids, left partially ajar and monitored, or replaced on ventilation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9536826PMC
http://dx.doi.org/10.30802/AALAS-JAALAS-22-000028DOI Listing

Publication Analysis

Top Keywords

ivc sealed
16
sealed lids
16
mice removed
12
ppm
12
ivc
9
lids
9
mouse rat
8
rat disposable
8
ivc removed
8
mechanical ventilation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!