Serous body fluid evaluation using the new automated haematology analyser Mindray BC-6800Plus.

Clin Chem Lab Med

Haematology and Cytology Unit. CORE Laboratory. Biochemistry and Molecular Genetics Department, Biomedical Diagnostic Centre, Hospital Clinic of Barcelona-IDIBAPS, Barcelona, Spain.

Published: October 2022

AI Article Synopsis

Article Abstract

Objectives: Cellular analysis of body fluids (BF) has clinical relevance in several medical conditions. The objective of this study is twofold: (1) evaluate the analytical performance of the BF mode of Mindray BC-6800 Plus compared to manual counts under microscopy and (2) analyse if the high-fluorescent cell counts provided by the analyser (HF-BF) are useful to detect malignancy.

Methods: A total of 285 BF was analysed: 250 corresponding to patients without neoplasia and 35 to patients with malignant diseases. Manual differential counts were performed in BF with ≥250 cells/μL. Percentages and absolute counts were obtained on the BC-6800Plus for total nucleated cells (TC-BF), mononuclear, polymorphonuclear and HF-BF. Statistical analysis was performed using Mann-Whitney U-test, Spearman's correlation, Passing-Bablok regression, Bland-Altman graph and ROC curve.

Results: To compare manual and automatic total cell counts, samples were divided in three groups: <250, 250-1,000 and >1,000 cells/μL. Correlation was good in all cases (r=0.72, 0.73 and 0.92, respectively) without significant differences between both methods (p=0.65, 0.39 and 0.30, respectively). The concordance between methods showed values of 90%. Considering malignant samples, median HF-BF values showed significant higher values (102 cells/μL) with respect to non-malignant (4 cells/μL) (p<0.001). The cut-off value of 8.5 HF-BF/μL was able to discriminate samples containing malignant cells showing sensitivity and specificity values of 89 and 71%, respectively. Considering both, HF-BF and TC-BF values, sensitivity and specificity values were 100 and 53%, respectively.

Conclusions: This study reveals that the Mindray BC-6800Plus offers an accurate and acceptable performance, showing results consistent with the manual method. It is recommended to consider both HF-BF and TC-BF values for the screening of the microscopic evaluation to ensure the detection of all malignant samples.

Download full-text PDF

Source
http://dx.doi.org/10.1515/cclm-2022-0447DOI Listing

Publication Analysis

Top Keywords

cell counts
8
counts
5
serous body
4
body fluid
4
fluid evaluation
4
evaluation automated
4
automated haematology
4
haematology analyser
4
analyser mindray
4
mindray bc-6800plus
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!